16
4. OPERATION PROCEDURES
Defrost
1. Remove chamber
Remove the flask mounting vacuum chamber from
the trap.
If ice has formed on trap wall, turn ON the defrost
switch.
2. Remove ice from trap
When ice has melted sufficiently to remove, do so
by pulling up on handle and turn OFF the defrost
switch.
Defrost auto stop.
DC401: Approx. 20min.
Turn OFF defrost switch manually.
3. Drain water
Remove the drain hose cap, and drain remaining
water from the trap into a drain basin or other
receptacle.
4. Stop refrigeration system
DEFROST
ON
OFF
ON
OFF
REFRIGE
OFF
1. Turn "OFF" the refrigeration switch.
2. Turn "OFF" the main power switch (ELB).
Flask Vacuum
Chamber
(Optional)
Drain Hose
Handle
Ice
Summary of Contents for DC401
Page 2: ......