11
Alignment
Introduction and Precautions
The
VX-7R
has been carefully aligned at the factory for
the specified performance across the specified amateur
bands. Realignment should therefore not be necessary
except in the event of a component failure. All compo-
nent replacement and service should be performed only
by an authorized VERTEX STANDARD representative,
or the warranty policy may be voided.
The following procedures cover the sometimes critical
and tedious adjustments that are not normally required
once the transceiver has left the factory. However, if dam-
age occurs and some parts are replaced, realignment may
be required. If a sudden problem occurs during normal
operation, it is likely due to component failure; realign-
ment should not be done until after the faulty component
has been replaced.
We recommend that servicing be performed only by
authorized VERTEX STANDARD service technicians,
who are experienced with the circuitry and fully equipped
for repair and alignment. Therefore, if a fault is suspect-
ed, contact the dealer from whom the transceiver was
purchased for instructions regarding repair. Authorized
VERTEX STANDARD service technicians realign all cir-
cuits and make complete performance checks to ensure
compliance with factory specifications after replacing any
faulty components.
Those who do undertake any of the following align-
ments are cautioned to proceed at their own risk. Prob-
lems caused by unauthorized attempts at realignment are
not covered by the warranty policy. Also, VERTEX STAN-
DARD must reserve the right to change circuits and align-
ment procedures in the interest of improved performance,
without notifying owners.
Under no circumstances should any alignment be at-
tempted unless the normal function and operation of the
transceiver are clearly understood, the cause of the mal-
function has been clearly pinpointed and any faulty com-
ponents replaced, and the need for realignment deter-
mined to be absolutely necessary.
Required Test Equipment
H
RF Signal Generator with calibrated output level at 500 MHz
H
Deviation Meter (linear detector)
H
In-line Wattmeter with 5% accuracy at 500 MHz
H
50-ohm, 10-W RF Dummy Load
H
8-ohm AF Dummy Load
H
Regulated DC Power Supply adjustable from 3 to 15 V DC, 3A
H
Frequency Counter: 0.2-ppm accuracy at 500 MHz
H
AF Signal Generator
H
AC Voltmeter
H
DC Voltmeter: high impedance
H
UHF Sampling Coupler
H
SINAD Meter
Alignment Preparation & Precautions
A 10-W RF dummy load and in-line wattmeter must be
connected to the main antenna jack in all procedures that
call for transmission, alignment is not possible with an
antenna. After completing one step, read the next step to
see if the same test equipment is required. If not, remove
the test equipment (except dummy load and wattmeter,
if connected) before proceeding.
Correct alignment requires that the ambient tempera-
ture be the same as that of the transceiver and test equip-
ment, and that this temperature be held constant between
68~86°F (20~30°C). When the transceiver is brought into
the shop from hot or cold air, it should be allowed some
time to come to room temperature before alignment.
Whenever possible, alignments should be made with os-
cillator shields and circuit boards firmly affixed in place.
Also, the test equipment must be thoroughly warmed up
before beginning.
Note: Signal levels in dB referred to in the alignment procedure
are based on 0 dBµ=0.5 µV (closed circuit).
Alignment Setup
Wattmeter
In-Line
RF Load
50-ohm
Coupler
Sampling
Meter
Deviation
Counter
Frequency
Generator
RF Signal
Meter
SINAD
AF Load
8-ohm
DW
Summary of Contents for VX-7R
Page 5: ...5 Block Diagram ...
Page 6: ...6 Interconnection Diagram ...
Page 21: ...21 RF Unit Lot 4 Circuit Diagram ...
Page 35: ...35 AF Unit Lot 3 Circuit Diagram ...
Page 37: ...37 AF Unit Lot 4 Circuit Diagram ...
Page 47: ...47 CNTL Unit Lot 3 Circuit Diagram ...
Page 49: ...49 CNTL Unit Lot 4 Circuit Diagram ...
Page 58: ...58 CNTL Unit Note ...
Page 63: ...63 Note ...