4.2.6 Calibration history
Calibration history
Fig. 4-1
Calibration historySensoLyt
®
700 IQ H
The calibration history provides the following information:
Date of the calibration
Slope [mV/pH]
Asymmetry potential [mV]
Calibration evaluation:
–
o.k.
: Successful calibration.
The new calibration data are taken over for measurement.
–
Error
: Calibration unsuccessful.
Sensor blocked for measurement.
4.2.7 Reactivating a valid calibration
The SensoLyt
®
700 IQ H provides a feature with witch you can reacti-
vate the last valid calibration if necessary. Thus you can immediately
continue to measure if a calibration failed.
Reactivating the
calibration data
Reactivating old calibration data is a temporary measure.
Take into consideration that the sensor may provide wrong
measured values. Ensure the correct functioning of the
sensor by checking and/or recalibrating it.
Currently active
calibration
Chronological list of
the last calibration
procedures
1
Open the setting table (see section
2
In the
Calibration
menu, select the setting
last valid
and then
exit the setting table with
Save and quit
.