VARiON
®
Plus
700 IQ H
Maintenance and electrode exchange
39
ba76207e02
12/2020
6
Maintenance and electrode exchange
6.1
General maintenance instructions
Maintenance condition
We recommend to switch on the maintenance condition each time the
sensor is taken out of the measuring position. This avoids unwanted
reactions of linked outputs. More detailed information on the mainte-
nance condition is given in the respective IQ
S
ENSOR
N
ET
system oper-
ating manual.
VARiON
®
/Epack
The VARiON
®
/Epack set with usual replacement parts is available for
maintenance (see chapter 7 R
6.2
Exterior cleaning
With normal operation (e.g. municipal wastewater) we strongly recom-
mend to clean the outside of the sensor:
when it is strongly contaminated (after visual check)
if erroneous measured values are suspected
each time before removing or exchanging an electrode
WARNING
Chemical or biological hazard.
Contact with the sample can be harmful to the user.
Depending on the type of sample, suitable protective
measures must be taken (protective clothing, gog-
gles, etc.).
To keep the electrodes clean, we recommend to use the
CH cleaning head (see chapter 7 R
).
NOTICE
Do not use any detergent or deionized water for clean-
ing. Heed the notes in section 3.2.1 E
Summary of Contents for YSI VARiON Plus 700 IQ H
Page 60: ...Contact Information VARiON Plus 700 IQ H 60 ba76207d02 12 2020 ...