VARiON
®
Plus
700 IQ H
Commissioning
13
ba76207e02
12/2020
Cleaning with washing-up liquid (tenside!)
Cleaning with or storing in deionized water
Storing in wrong storing solution
Therefore, exactly follow the instructions in the two following chapters.
3.2.2 Reference electrode
Commissioning
Fig. 3-1
Reference electrode with storing aids
In the delivery condition, the electrode is equipped with a watering cap
and a nut that protects the screw-in thread. The watering cap contains
3 mol/l potassium chloride solution. Before mounting, unscrew the
watering cap. Then, using the special hexagon key, unscrew the elec-
trode from the nut. Keep both storing aids in case you want to store the
electrode.
Notes on storage
If you will not use the electrode for a longer period of time, screw the
electrode into the nut as far as it will go. Fill the watering cap to the brim
with 3 mol/l potassium chloride solution and tightly screw the watering
cap on the electrode by hand.
The junction of the reference electrode must not
dry up (follow notes on storage)
be damaged
be brought into contact with grease.
Watering cap
Junction
Reference electrode
Nut
Sealing ring