51
Calibration
ISE Calibration 3-Point
1.
“Calibration setup (pH, ORP, ISE, conductivity, turbidity,
and total algae sensors)” on page 33
. Rinse the sensor 2-3 times
with a small amount of standard #1.
NOTE:
It is best to calibrate in order of increasing concentration
(e.g. if using 1 mg/L and 100 mg/L standards, calibrate with
1 mg/L first).
2.
Push the
Cal
key, then select the applicable ISE sensor.
3.
Carefully immerse the sensors into a solution of standard #1.
4.
Allow the temperature of the standard to stabilize, then select
Calibration value
. Enter the calibration value that corresponds to
standard #1.
5.
Observe the actual measurement readings for stability (white line
on graph shows no significant change for 40 seconds), then select
Accept Calibration
(
Figure 59
). “Ready for cal point 2” will be
displayed in the message area.
6.
Select
Finish Calibration
to complete a 1-point calibration.
Otherwise, continue the calibration procedure to complete at least
a 2-point calibration.
NOTE:
A 2-point calibration is extremely accurate and is the
preferred method.
7.
Rinse the sensor 2-3 times with a small amount of standard #2.
Discard the standard after rinsing.
8.
Carefully immerse the sensors into a fresh solution of standard #2.
9.
Allow the temperature of the solution to stabilize then select
Calibration value
. Enter the calibration value that corresponds to
standard #2.
10.
Observe the actual measurement readings for stability (white line
on graph shows no significant change for 40 seconds), then select
Accept Calibration
(
Figure 59
). “Ready for cal point 3” will be
displayed in the message area.
11.
Select
Finish Calibration
to complete a 2-point calibration.
Otherwise, continue the calibration procedure to complete a
3-point calibration.
NOTE:
To calibrate with a chilled third standard, see
.
12.
Rinse the sensor 2-3 times with a small amount of standard #3.
Discard the standard after rinsing.
13.
Carefully immerse the sensors into a fresh solution of standard #3.
14.
Allow the temperature of the solution to stabilize then select
Calibration value
. Enter the calibration value that corresponds to
standard #3.
15.
Observe the actual measurement readings for stability (white line
on graph shows no significant change for 40 seconds), then select
Accept Calibration
. “Calibration successful!” will be displayed in
the message area.
Figure 59
Calibrate ISE
Summary of Contents for YSI ProDSS
Page 1: ...USER MANUAL ProDSS USER MANUAL DOCUMENT 626973 01REF...
Page 2: ...i This page left intentionally blank...
Page 58: ...56 Calibration This page left intentionally blank...
Page 74: ...72 Maintenance and Storage This page left intentionally blank...
Page 90: ...88 Appendix A DO Calibration Values This page left intentionally blank...