IOEM/REM Safey and Installation Guide
76
IOEM/REM Safey and Installation Guide
77
Configure 5X00 system for IOEM 4/20mA output
1.
Follow steps 1-6 of hardware installation-starting on page 25;
2.
Wire IOEM I/O for 4/20 output type- page 47;
3.
Verify all steps at Using AquaManager for IOEM configuration - page
73 are completed;
4.
Open AquaManager>open 5X00 node window>
;
5.
Select 5X00 system tab to be configured;
6.
Configure 5X00 4/20mA output system fields;
5X00 Systems Management PID control mode with 4/20mA output configuration
Notes:
-
must be selected to make 4/20mA fields live;
-
Proper wiring of IOEM output channel is required in order for IOEM
4/20mA output device to work correctly.
-
IOEM Source channels are only available at 5X00 system fields once
IOEM mode has been uploaded using AquaManager-page 73;
-
IOEM Sources will only be available at IOEM and channel mode pull-
down menus that have been configured as a 4/20mA output type;
For 5X00
system configuration field information see - AquaManager Help>Contents
tab>Working with Nodes>Systems management>Systems Management
fields table and link to specific sensor system being configured bookmarks;
Configuring IOEM/REM expansion modules
-
4/20mA output channels can only be assigned to 1 5X00 system at a
time
. When selecting a 4/20mA output channel that is currently pending
or confirmed AquaManager displays the below 4/20mA Conflict Error
window.
will not be live until the 4/20mA assignment
conflict is resolved
-
4/20mA output values cannot be viewed at 5X00 instruments or at
AquaManager.
7.
to upload system configuration to networked instrument;
8.
Test configured output functionality after reconfiguring -
see
Chapter 4
5X00 instrument manuals Configuring>Sensor Setup>Service> Output test
system menu to verify 4/20mA output.
9.
Verify 4/20mA signal.
The 4/20mA output signal is not displayed at the
Run Screen. Therefore, it is strongly recommended to verify the 4/20mA
signal at the IOEM or third party device.
Configure 5X00 system for IOEM PID output
control device
1.
Follow steps 1-6 of hardware installation-starting on page 25;
2.
Wire IOEM I/O for 4mA-20mA output type- page 47;
3.
Verify all steps at Using AquaManager for IOEM configuration - page
73 are completed;
4.
Open AquaManager>open 5X00 node window>
;
5.
Select 5X00 system tab to be configured;
6.
Configure 5X00 system fields
-
For 5X00 system configuration field information see - AquaManager
Help>Contents tab>Working with Nodes>Systems management>Systems
Management fields table and link to specific sensor system being con-
figured bookmarks;
Configuring IOEM/REM expansion modules