The same .mdb file is used for all 5X00 and IOEM/REM instruments.
Once the upload process begins the instrument type is determined and
the appropriate firmware is uploaded.
7.
at Firmware Updater window. Firmware Update info window
opens;
Installation and Wiring
8.
Left click “Yes” to begin firmware upload. Firmware Updater screen opens;
9.
at
window when upload
completes;
10. Verify uploaded version information is correct at IOEM/REM node after
manually polling IOEM/REM instruments.
When all instruments have
been uploaded with new firmware, it is important to open and poll all nodes
to ensure latest firmware version is stored at AquaManager database. After
manual poll,
it may be necessary to close node window at AquaManager and
reopen in order for latest firmware version to be displayed at node window.