FDO 4410(W)

Maintenance, cleaning, replacement

11

ba76197e01

10/2017

3.2

Exchanging the

 

sensor cap

To exchange the sensor cap, proceed as follows:

Removing the

sensor cap

Mounting the

sensor cap

Before removing the sensor cap, observe the general mainte-
nance instructions in section 3.1.

1.

Pull the sensor out of the sample.

2.

Clean the outside of the sensor (see section 3.3.1).

3.

Unscrew the protective hood from the sensor by hand. 

4.

Thoroughly clean and dry the sensor head once again.

5.

Grasp the sensor cap on the sides (arrows in figure on the left) and 
remove it by pulling it away from the sensor in a straight upward 
direction.

Caution:
Do not push any tools or other sharp objects between the sealing sur-
faces. This would damage the sealing surfaces and sealing ring.

6.

Check the front surface of the sensor for absolute cleanness and 
clean it if necessary (see section 3.3.2).

7.

Thoroughly clean the thread of the protective hood.

8.

Check the sealing ring for intactness and correct position.
The sealing ring has to be replaced if it is damaged.

9.

Aim the sensor head upwards and place the new sensor cap on the 
sensor. The locking device on the sensor head must be inserted into 
the receptacle on the inside of the sensor cap (see figure on the left).

10.

Put the protective hood on the sensor head and screw it tight by hand 
as far as it will go. A gap of approx. 0.8 mm remains between the pro-
tective hood and sensor.
The sensor is immediately ready to measure.

Sealing
ring

Summary of Contents for YSI FDO 4410

Page 1: ...OPERATIONS MANUAL ba76197e01 10 2017 FDO 4410 FDO 4410W OPTICAL D O SENSOR FDO 4410 FDO 4410W OPERATING MANUAL...

Page 2: ...2 ba76197e01 10 2017 FDO 4410 Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email environmental ysi com Internet www ysi com Copyright 2017 Xylem Inc...

Page 3: ...pening and closing the IDS plug connection FDO 4410W 7 2 4 Measuring 8 2 5 Function check and user calibration 8 2 6 Storage 9 3 Maintenance cleaning replacement 10 3 1 General maintenance instruction...

Page 4: ...tronics 3 Connection cable FDO 4410 4 IDS plug FDO 4410W 5 Protective hood 6 Sensor membrane 7 SC FDO 4410 sensor cap with memory chip 8 Gold plated contacts for the memory chip of the sensor cap 9 Lo...

Page 5: ...ensor is connected and is used for measurement and for measured value documentation The calibration data are stored in the sensor cap Thus the calibration is automatically retained if the sensor or me...

Page 6: ...ure scratches 2 2 Commissioning Scope of delivery D O sensor FDO 4410 W with sensor cap Check and storage vessel FDO Check Operating manual Commissioning Connect the electrode to the meter FDO 4410 vi...

Page 7: ...ng device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plug connecti...

Page 8: ...d application see page 5 the measuring characteristics of the sensor cap remain stable for the specified service life Thus a user calibration is not usually required When does a check or user calibrat...

Page 9: ...for the temperatures of the sensor and cali bration vessel to adjust 2 6 Storage Always store the sensor in the check and storage vessel FDO Check at a temperature in the range 0 50 C 32 122 F The ste...

Page 10: ...clean and dry prior to removing the sensor cap Please do not touch the outer sensor membrane with your fingers Touch the sensor cap at the sides only arrows in figure on the left Avoid any great mech...

Page 11: ...Do not push any tools or other sharp objects between the sealing sur faces This would damage the sealing surfaces and sealing ring 6 Check the front surface of the sensor for absolute cleanness and c...

Page 12: ...h In the case of persisting dirt you can add some household washing up liq uid to the tapwater Attention Never use any alcohol for cleaning 3 3 2 Interior cleaning of sensor cap and sensor head If moi...

Page 13: ...check the zero point of the sensor Measurement in a nitrogen atmosphere recommended method Measurement in a sodium sulfite solution according to DIN EN 25814 ISO 5814 Test criterion The sensor is OK...

Page 14: ...n 3 2 Dirt inside the sensor cap and sensor head Protective hood not properly tightened Sensor cap untight or defective Dismantle the sensor cap Clean the insides of the sensor cap and sensor head see...

Page 15: ...25 C 77 F Dimensions in mm Weights FDO 4410 180 g with 3 m cable FDO 4410W 107 g FDO Check 58 g Materials Shaft POM Connection head POM Sensor cap PVC silicone and PMMA Sensor head POM PV and PMMA Se...

Page 16: ...istance Sensor with connection cable IP 68 1 x 106 Pa or 10 bar Cable plug FDO 4410 IP 67 when plugged in The FDO 4410 W meets the requirements according to article 3 3 of the 97 23 EC directive Press...

Page 17: ...bar O2 partial pressure Response time at 20 C 68 F in stirred solution t90 90 of the final value display after 30 s t95 95 of the final value display after 45 s t99 99 of the final value display after...

Page 18: ...sensor cap SC FDO 4410 201 310Y Connection cable FDO 4410 W meter Description Model Order no IDS connection cable 1 5 m IDS CABLE 1 5 903 850Y IDS connection cable 3 m IDS CABLE 3 903 851Y Radio conn...

Page 19: ...uct for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Product Return for...

Page 20: ...Contact Information FDO 4410 20 ba76197d01 10 2017...

Page 21: ......

Page 22: ...and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with cus...

Reviews: