12
Setup
3.2
Setup the EXO Handheld Display
In the following section we will cover setting up EXO PAR utilizing the EXO Handheld Display.
1
. You will first need to enable the PAR(s) sensor in the EXO Handheld. To do so you will need to select
Handheld
,
and then select
Display
, and from the list of
Units
select PAR and you will see figure 13. On the next screen you
can enable either the PAR sensor or sensors by clicking the boxes shown, figure 14.
Figure 13
Select PAR from the list
of Units
Figure 14
Enable sensors
2
. Using the Calibration Certificate supplied with your Li-Cor PAR sensor, you will now enter the multipliers into EXO
Handheld Display. To do so first select
Calibration
and from the list of sensors displayed select PAR sensor. If
successful you will see the screen shown in figure 15. From the Calibration Certificate you will use the In-Water
multiplier that is listed under the Handheld Meters and Loggers section. You will enter the multipliers exactly as
they are on the certificates in the fields shown for Channel 1 and Channel 2 shown in figure 16. If you are only
using one PAR sensor, the default channel will be Channel 1. Leave the default value in Channel 2, which will be (1).