Xylem WTW OxiTop-IDS Operating Manual Download Page 29

Multi 3630/3620 IDS

Operation with OxiTop

®

-IDS measuring heads

ba77045e01

09/2019

29

5.5

Calibration

5.5.1 Why calibrate?

OxiTop

®

-IDS measuring heads age. This changes e.g. the tightness of the Oxi-

Top

®

-IDS measuring heads and the accuracy of the pressure measurement. 

The condition of the measuring heads can be evaluated by checking their tight-
ness and measuring accuracy (see BOD application report).

5.5.2 When do you have to calibrate?

Routinely within the framework of the company quality assurance

5.5.3 Calibration procedure

Creation of a defined negative pressure and balancing with the display (see 
BOD application report).
In contrast to physical or chemical measuring methods, BOD systems do not 
provide standardized values, as the use of microorganisms has a great spec-
trum depending on their composition and food supply. With the aid of testing 
tools and suitable tests, errors due to defective pressure sensors or untight 
OxiTop

®

-IDS measuring head threads can be identified.

You can check the OxiTop

®

-IDS measuring heads with the following proce-

dures.

OxiTop

®

-PM (Cal-Test)

OxiTop

®

-PT (pneumatic test)

Key

Explanation

<PRT_>

Output of the data as a *.csv file to the USB-A interface, to a 
connected USB memory device.

<PRT>

Output of the data in ASCII or CSV format (setting in the system 
menu) to the interface USB-A or USB-B.
The meter recognizes whether a printer or PC is connected.

The data transferred to the USB memory device are in the directory 
"OxiTop".

Details on the tests are given in the operating manual of your 
OxiTop

®

-IDS measuring head.

The test OxiTop

®

-PT (pneumatic test) is operated exclusively at the 

OxiTop

®

-IDS measuring head. The meter is not required for this.

Summary of Contents for WTW OxiTop-IDS

Page 1: ...OPERATING MANUAL ba77045e01 09 2019 OxiTop IDS B Multi 3630 3620 IDS WIRELESS OPERATION OF OXITOP IDS B MEASURING HEADS 3620 IDS...

Page 2: ...Multi 3630 3620 IDS 2 ba77045e01 09 2019 Copyright 2019 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...management 11 4 2 1 Overview 11 4 2 2 Overview of the sets OxiTop set management 12 4 2 3 Overview of the samples of the set 13 4 2 4 Overview of the OxiTop IDS measuring heads in a sample 15 4 3 Sett...

Page 4: ...Contents Multi 3630 3620 IDS 4 ba77045e01 09 2019 6 Technical data 31 7 What to do if 32 8 Firmware update for OxiTop IDS measuring heads 33...

Page 5: ...requisites Meter Multi 3630 IDS or Multi 3620 IDS with current software version software update possible Adapter WLM M for wireless connection to IDS sensors OxiTop IDS B measuring heads Data exchange...

Page 6: ...might be damaged if the actions mentioned are not taken 2 1 2 Safety signs on the meter Note all labels information signs and safety symbols on the meter and in the battery compartment A warning symb...

Page 7: ...safe operation Note the following points for safe operation The meter may only be operated according to the authorized use specified above The meter may only be supplied with power by the energy sour...

Page 8: ...20 IDS meter see operating manual of the Multi 3630 3620 IDS meter For initial commissioning press M_ to switch the meter to operation with OxiTop IDS measuring heads Switch on the OxiTop IDS measurin...

Page 9: ...DS measuring heads OxiTop IDS B measuring heads Undisturbed radio link between OxiTop IDS measuring head and meter The radio connection is active at the OxiTop IDS measuring head No IDS sensor is conn...

Page 10: ...g ing The AutoTemp function can be switched on BOD standard The measuring range and sample volume can be selected from seven fixed combi nations The incubation temperature is 20 C BOD special Measurin...

Page 11: ...ode The meter automatically creates an identifiable name for the sample The name consists of a letter A and an index x that is automatically increased for fur ther samples in the same set e g A 1 The...

Page 12: ...guage etc 1 Set name Check mark finished when the data of the measuring head have been transmitted to the meter at the end of the specified Duration 2 Status of the set Number of measured values maxim...

Page 13: ...sured value if the Warning difference pressure is exceeded in at least one OxiTop IDS measuring head of the sample only for Measuring mode Pressure p 3 Status line number of OxiTop IDS measuring heads...

Page 14: ...Operation Multi 3630 3620 IDS 14 ba77045e01 09 2019...

Page 15: ...transmitted to the meter at the end of the specified Duration 3 Series number of the OxiTop IDS measuring head 4 Measured value last value transmitted by the measuring head in the specified interval...

Page 16: ...AutoTemp on off Function to control the temperature adjustment Measurement is started automatically at least 1 hour at most 3 hours after the OxiTop IDS measuring head is started Pre tempering the sa...

Page 17: ...akes place LED on off Switching on off the LED indication of the operating statuses at the OxiTop IDS measuring head The LED at the OxiTop IDS measur ing head can be switched off to save energy Number...

Page 18: ...standard 0 40 80 200 mg l 0 400 mg l 0 800 2000 4000 mg l The measuring range expected for the sample can be selected from 7 perma nent ranges Selecting the measuring range defines the Sample volume T...

Page 19: ...ermostat cabinet box BOD standard temperature 20 C 1 C Brown sample bottles bottle volume 510 ml Magnetic stirring rods 1 stirring rod for each sample bottle Magnetic stirring rod remover Suitable ove...

Page 20: ...bottles that are not damaged e g are not cracked chipped or scratched Determine the reaction conditions temperature quantity under which no dangerous excess pressure develops by means of monitored pre...

Page 21: ...iew of the sets started is dis played During the initial commissioning the list OxiTop set management is empty The display shows No data available 4 In the OxiTop set management use F1 New to create a...

Page 22: ...y preselected as name for the set The name can be changed freely max 8 characters numerals letters Duration Settings 7 Open the Settings menu 8 Check and if necessary change the settings for the set i...

Page 23: ...als letters Measuring range only with Type Standard Calculation only with Type BOD special Settings only with Type BOD special 12 Further information and settings of the sample are displayed as an Inf...

Page 24: ...en the OxiTop IDS measuring head has been identified by the meter the following data of the head are displayed ID assigned at the measuring head Series number of the OxiTop IDS measuring head 1 Number...

Page 25: ...ach OxiTop IDS measuring head connected to the meter can indicate the following data on the display name of the set name of the sample settings For this press the ENTER key at the measuring head 21 Ke...

Page 26: ...measuring head with many measurement data it may take longer In the OxiTop set management the set is given a checkmark in the status field 26 If necessary edit the evaluation of the sample e g if an...

Page 27: ...n the details of the selected sample with MENU ENTER or F1 i Using MENU ENTER display the individual measuring heads of the selected sample There you can view the data of the individual measuring head...

Page 28: ...aces and on the settings for the USB interfaces is given in the operating man ual of your meter 1 Connect the output device to a USB interface USB A USB printer USB memory device USB B PC 2 Using M op...

Page 29: ...e use of microorganisms has a great spec trum depending on their composition and food supply With the aid of testing tools and suitable tests errors due to defective pressure sensors or untight OxiTop...

Page 30: ...sample bot tle The reduction of oxygen causes a defined negative pressure The BOD value corresponding to this negative pressure is given in the packet The procedure is similar to a BOD5 measurement w...

Page 31: ...S IDS WLM M adapter OxiTop IDS measuring head Operation with OxiTop IDS measuring heads Measuring ranges Measuring mode BOD standard 0 40 80 200 400 800 2000 4000 mg l BOD Measuring mode BOD special 0...

Page 32: ...aluation Temper the sample more precisely See error symptom A OxiTop IDS measuring head cannot be addressed An measuring head cannot be addressed The radio link was switched off at the OxiTop IDS meas...

Page 33: ...e Multi 3630 3620 IDS operating manual 1 Install the downloaded firmware update on a PC An update folder is created in the Windows start menu If an update folder for the OxiTop IDS measuring head is a...

Page 34: ...Firmware update for OxiTop IDS measuring heads Multi 3630 3620 IDS 34 ba77045e01 09 2019...

Page 35: ......

Page 36: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: