Multi 3630/3620 IDS
Operation with OxiTop
®
-IDS measuring heads
ba77045e01
09/2019
21
5.1.2 The course of the measurement
Preparation before
starting the
measurement
1
Carry out the preparatory activities
(see section 5.1.1 P
Keep ready the instruments and accessories required
Determining the sample volume
Preparing the samples
2
Switch on the meter with
<On/Off>
.
3
If necessary, use
<M_>
to switch to operation with the OxiTop
®
-IDS
measuring heads.
The
OxiTop set management
(overview of the sets started) is dis-
played.
During the initial commissioning the list
OxiTop set management
is
empty.
The display shows
No data available!
.
4
In the
OxiTop set management
, use
<F1>
/[
New
] to create a new set of
samples.
The
Set
menu opens.
5
Adjust the temperature of the solutions and measure the current tem-
perature if the measurement is made without a temperature sensor.