Multi 3630 IDS
Settings
ba77170e03
10/2017
87
11.6.3 Automatic
Stability control
The automatic
Stability control
(AutoRead) function continuously checks the
stability of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
You can activate or switch off the automatic
Stability control
function (see sec-
The measured parameter flashes on the display
as soon as the measured value is outside the stability range
when you switch over between the measured parameters with
<M>
.
when the automatic
Stability control
is switched off.
11.6.4 Automatic switch-off function
The instrument has an automatic switchoff function in order to save the batter-
ies (see section 11.6.1). The automatic switch-off function switches off the
meter if no key is pressed for an adjustable period.
The automatic switch-off function is not active
if the power pack is connected
if a USB-B cable is connected
if the
Automatic data storage
function is active, or with automatic data trans-
mission
11.6.5 Display illumination
The meter automatically switches the display illumination to energy saving
mode if no key is pressed for 20 seconds.
The illumination is switched on with the next keystroke again.
You can also generally switch on the display illumination (see section 11.6.1).
11.7 Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
Detailed information on the memory functions of the
Multi 3630 IDS is given in section 12.