Operation
Oxi 3310
40
ba75804e08
10/2017
Starting the automatic
storing function
To start the automatic storing function, select
Continue
with
<
><
>
and confirm with
<ENTER>
. The meter switches to the measured value
display.
The active automatic data storing function can be recognized by the
progress bar in the status line. The progress bar indicates the remain-
ing storing duration.
Note
If the automatic storing function is activated, only the following keys are
active: Softkeys,
<M>
,
<STO
_
>
and
<On/Off>
. The other keys and
the automatic switch-off function are deactivated.
Energy saving mode
([Eco mode])
If the automatic storing function is active, the meter provides an energy
saving mode ([Eco mode]) to avoid unnecessary energy consumption.
The energy saving mode switches off functions of the meter that are not
required for the automatic storing of measurement data (such as the
display). By pressing any key the energy saving mode is switched off
again.
Duration
1 min
... x min
Storing duration.
Specifies after which time
the automatic data storing
should be terminated.
The lower limit of the stor-
ing duration is restricted by
the storing interval.
The upper limit is restricted
by the number of free
memory locations.
Menu item
Possible
setting
Explanation
O
2
7.92
mg/l
25.0
°C
0d03h00min
[SAL]
15.03.2014
08:00
Remaining storing duration
Graphical display of the
storing duration
Eco mode
Summary of Contents for WTW Oxi 3310
Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Page 63: ......