Operation
Oxi 3310
32
ba75804e08
10/2017
the instrument.
When to calibrate?
After connecting another D.O. sensor
When the sensor symbol flashes (after the calibration interval has
expired).
Calibration datasets
The Oxi 3310 administrates two sets of calibration data:
Sensors of different types can be calibrated separately from each other.
When one sensor type is calibrated, the calibration data of the other
type remain stored. The Oxi 3310 recognizes the type of the connected
sensor and automatically uses the correct calibration data.
Calibration procedures
The Oxi 3310 provides 2 calibration procedures:
Calibration in water vapor-saturated air.
Use an OxiCal
®
air calibration vessel for the calibration.
Calibration via a comparison measurement (e.g. Winkler titration
according to DIN EN 25813 or ISO 5813). At the same time, the rel-
ative slope is adapted to the comparison measurement by a correc-
tion factor. When the correction multiplier is active, the
[Factor]
indicator appears in the measuring window.
Stability control
(AutoRead)
In calibration, the Stability control function (AutoRead) is automatically
activated.
Display calibration data
and output to interface
You can have the data of the last calibration displayed (see section
4.4.5). Subsequently, you can transmit the displayed calibration data to
the interface, e.g. to a PC, with the
<F2>
/
[USB output]
key.
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Sample record
Calibration evaluation
After calibration, the meter automatically evaluates the current status of
Set 1 for the type, "CellOx":
– CellOx 325
Set 2 for the type, "DurOx":
– DurOx 325
Oxi 3310
Ser. no. 08502113
CALIBRATION
O
2
15.03.2014 08:00:33
DurOx 325
Relative slope
0.88
Temperature
25.0 °C
++
etc...
Summary of Contents for WTW Oxi 3310
Page 1: ...OPERATING MANUAL ba75804e08 10 2017 Oxi 3310 DISSOLVED OXYGEN D O METER...
Page 2: ...2 ba75804e08 10 2017 Oxi 3310 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...
Page 5: ...Oxi 3310 Contents 5 ba75804e08 10 2017...
Page 10: ...Overview Oxi 3310 10 ba75804e08 10 2017...
Page 14: ...Safety Oxi 3310 14 ba75804e08 10 2017...
Page 48: ...Operation Oxi 3310 48 ba75804e08 10 2017...
Page 51: ...Oxi 3310 Maintenance cleaning disposal 51 ba75804e08 10 2017...
Page 53: ...Oxi 3310 What to do if 53 ba75804e08 10 2017...
Page 56: ...Technical data Oxi 3310 56 ba75804e08 10 2017...
Page 60: ...Lists Oxi 3310 60 ba75804e08 10 2017...
Page 62: ...Firmware update Oxi 3310 62 ba75804e08 10 2017...
Page 63: ......