Multi 9430
Ion concentration
ba75966e04
04/2015
57
6.3.1
Standard addition
In the "Standard addition" procedure, a known amount of standard solution is
added to the sample.
The ion concentration in the sample is calculated from the change in potential.
Measuring
1.
Select the measuring method (see section 6.3 S
2.
Prepare two standard calibration solutions.
3.
Perform a two-point calibration according to user guidance.
4.
The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.
01.08.2013 08:00
01.08.2013 08:00
#1 0.010 mg/l
20.2 mV 25.0 °C
#2 0.020 mg/l
79.2 mV 25.0 °C
Slp.: 54.1 mV
++ (Ion type: Ag)
CALIBRATION
ADA 94/IDS DIN
B092500013
5.
Start the measurement with
<F1>
/[
continue
].
An entry window appears.
6.
Prepare the sample.
7.
If necessary, add the ISA/TISAB solution to the sample.
8.
Thoroughly rinse the electrode with deionized water.
9.
Immerse the electrode in the sample.
10.
Using
<
> <
>
and
<ENTER>
, select the values for the volume of
the sample (
Sample volume
) and the volume of the ISA/TISAB solution
(
ISA/TISAB vol.
).
i
Immerse sensor in sample
Sample volume
100.0 ml
ISA/TISAB vol.
1.0 ml
continue
Standard addition
01.08.2013 08:00