Calibration record
(example)
4.12
Depth of immersion (MPP 9x0 IDS multi parameter
probe)
4.12.1 General information
The multi parameter probes of the MPP 9x0 IDS series and the respec-
tive IDS sensors of the XXX-P series enhance your Multi 3430 meter
by the display of the parameter, depth of immersion (DPT).
The parameter depth of immersion is integrated as a secondary mea-
sured value in the display of all main measured values.
Example: Depth of immersion as main and secondary measured value
(MPP 9x0 IDS with a pH-IDS sensor of the XXX-P series)
When the probe is equipped to capacity (MPP 930 with 3 IDS sensors
of the XXX-P series), you can simultaneously measure three main
parameters (pH, O2,
ϰ
) and two secondary parameters (temperature
and depth of immersion).
1
Depth of immersion as a main measured value
2
Depth of immersion as a secondary measured value
Multi 3430
Ser. no. 12345678
CALIBRATION TRB:
VisoTurb 900-P
Ser. no. 14E999003
18.09.2014 08:09:10
# 1
0.0 FNU
# 2
124.0 FNU
Sensor
+++
_____________________________________
1
2
30.09.2012 08:00