Technical data
K 800
Slope too low
Technical data
Measuring range
Reproducibility
pH range
Temperature range
Membrane resis-
tance
Length
Diameter
Cable length
Plug
DIN plug
– Junctions encrusted
– Leave the bridge electrolyte to react on
the junctions for some hours until the
crusts have dissolved.
– Cable broken
– Exchange the electrode
Cause
Remedy
Cause
Remedy
– Conditioning time too short
– Extend the conditioning time
– Standard solutions too old
– Use new standard solutions
– Junctions encrusted
– Leave the bridge electrolyte to react on
the inner junctions for some hours until
the crusts have dissolved.
– Electrode defective
– Exchange the electrode
0.04 ... 39,000 mg/l K
+
(1 x 10
-6
... 1 mol/L K
+
)
± 2 %
2 ... 12
(see I
)
0 ... 40 °C
< 50 M
Ω
170 mm (including 50 mm connection head)
Shaft:
12 mm
Connection head
16 mm
1 m