DIQ/S 281
Outputs
111
ba77244d01
08/2020
6.7.3
Fixed current value
Function
The
Fixed current value
function can be used to test the operativeness of the
instruments connected to the outputs: output different current values to the out-
put and, while doing so, check the behavior of the connected instrument.
Io
0 ... 20 mA
Current value on the output if
the measured value equals the
Nominal value
ti
0 ... 9999 s
Hold-back time:
Integral part of the controller
(
0
= not effective)
td
0 ... 9999 s
Reset time:
Differential part of the
controller
(
0
= not effective)
Behavior at error
Fixed current value
In the case of an error, the
current output supplies the
current value defined in the
Current with error
field (any in
the range
0 ... 21 mA).
Unchanged
In the case of an error, the
current on the output remains
unchanged.
Setting
Selection/Values
Explanation
Setting
Selection/Values
Explanation
Current output
Fixed current
value
Using
Save and quit
, the nominal
amperage that was input as
Inom
is output.
Inom
0 ... 20 mA
The nominal amperage that is
output.
The settings for other functions in the
Current output
menu as, for
example,
PID controller
and
Recorder
are retained while the
Fixed
current value
is carried out.