Technical data
Alyza IQ PO4
164
ba76201d03
05/2020
7.5
Consumption data
The consumption of chemicals depends on the specified intervals and on the se-
lected measuring range.
Typical
consumption
values
Cable (requirements)
Europe: IEC 60332-1-2
USA, Canada: UL 2556 VW-1
Details:
Temperature resistant
in the range -20 °C…+80 °C
(-68 °C…+176 °F),
Weather-resistant (year-round)
Watertight (cable sheath)
Copper wire
Power consumption
Alyza IQ
300 - 1600 W
(depending on the length of the heat tracing)
Basic consumption
300 W
Heat tracing, intake line
return line
16 W/m
(max. 80 m heat tracing allowed)
IQ S
ENSOR
N
ET
cable
(SNCIQ, SNCIQ/UG,
SACIQ)
Insulation
≥
500 V
Temperature resistant
in the range -20 °C…+80 °C
(-68 °C…+176 °F),
Weather-resistant (year-round)
Watertight (cable sheath)
Power delivery to the
IQ S
ENSOR
N
ET
max. 10 W
Solution
Sufficient for
Reagent solution
R-PO4/1-1A/B (MR1)
R-PO4/1-2A/B (MR2)
Examples:
– 3 months (measuring interval 5 min-
utes)
(only MR1)
– 6 months (measuring interval 10 min-
utes)
Cleaning solution
C-PO4/1-1
6 months (daily cleaning)
Standard solution
S-PO4/1-0.0/-1.1/-1.10/-40
6 months (daily calibration)
Summary of Contents for WTW Alyza IQ PO4