Tilt Raw Data
4.4.4
The
Tilt
variable displays the angle of the wading rod from the vertical based on one
second measured raw data.
a) The Tilt information is calculated from the internal pitch and roll sensor in the
FlowTracker2 probe.
b) The Pitch and Roll data is recorded once per second.
Battery Indicator
4.4.5
The
Battery
variable displays the real time battery voltage level of the batteries housed
in the FlowTracker2 handheld.
a) The FlowTracker2 uses eight AA batteries contained in a battery pack.
b) Cold weather can greatly affect battery voltage and capacity; always check battery
voltage after the system has acclimated to outside temperatures.
c) The Battery data are updated once per second.
d) Comparison of Battery types are listed in
Table 4-5 – Battery Specification
Category
Alkaline
NiMH
General description
Single use
Rechargeable
Nominal battery capacity (typical)
1500 mAh
1900mAh
New battery voltage
12V
11V
Drained battery voltage
8V
9V
When the remaining capacity of the battery reaches
0%
, the FlowTracker2 handheld
will supply the following message, “
Battery level critically low – shutting down
”. The
user will be required to replace the batteries before continuing with operations.
Pressure
(Pressure Sensor)
4.4.6
The
Pressure
variable displays the pressure measurements from the one second
measured raw data.
a) Pressure data are used for calculating the total water and point measurement
depth from the water surface.
b) Atmospheric or Air pressure calibration should be performed before and during a
measurement to compensate for changes in atmospheric pressure during a
measurement.
c) Pressure data are updated once per second.
Pressure data will not be displayed if the probe is in air. The probe must be submerged
in water. Pressure sensor data displayed in this function are NOT stored and enables
the user to verify the pressure sensor operations.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...