![Xylem SonTek FlowTracker2 User Manual Download Page 290](http://html1.mh-extra.com/html/xylem/sontek-flowtracker2/sontek-flowtracker2_user-manual_3888944290.webp)
SonTek – a Xylem brand
FlowTracker2 User’s Manual (January 2019)
289
Category
Value
Description
Logging Duration Per
File (min)
Defines a file logging limit based on a time
duration.
Logging Interval Start /
End Time Alignment
The Logging Interval Start / End Time
Alignment works in combination with
“Logging Duration Per File (min)”.
If your logging duration per file is specified
as 10 min and you start your logging at
17:03:25:
-
Without logging interval alignment
the next file will be started at
17:13:25, the third one at 17:23:25
and so on
-
With logging interval alignment the
next file will be started at 17:20:00,
the third one at 17:30:00 and so on
Table 15-2: Velocity Range Settings
Velocity
Range
Absolute max
velocity
in arbitrary
direction
in ideal
conditions
Absolute max
velocity
in X direction
in ideal
conditions
Recommended
max velocity
in arbitrary
direction
in ideal
conditions
Recommended
max velocity
in X direction
in ideal
conditions
Auto
± 112 cm/s
± 448 cm/s
± 60 cm/s
± 200 cm/s
± 10 cm/s
± 10 cm/s
± 40 cm/s
± 3 cm/s
± 10 cm/s
± 20 cm/s
± 14 cm/s
± 52 cm/s
± 7 cm/s
± 20 cm/s
± 50 cm/s
± 27 cm/s
± 108 cm/s
± 15 cm/s
± 50 cm/s
± 100 cm/s
± 55 cm/s
± 220 cm/s
± 30 cm/s
± 100 cm/s
± 200 cm/s
± 112 cm/s
± 448 cm/s
± 60 cm/s
± 200 cm/s
Ideal conditions
consist of:
-
ideal signal conditions (high SNR, low noise)
-
Ideal boundary conditions (no boundaries nearby)
-
Ideal temperature and sound speed (18°C and 1475 m/s)
-
For X direction velocity – uniform flow in X direction
2D/3D FT2 probe
would use the same recommendation with
Z velocity maximum
same as the arbitrary direction.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...