1.2. Definitions and Terminology
This section defines terms commonly used when working with FlowTracker2
•
Averaging Time
– The time (in seconds) in which the FlowTracker2 records data
at each measurement location. This is a user-specified value from 10 to 1000
seconds.
•
Handheld
– The FlowTracker2 is controlled from the keypad on the handheld.
The LCD screen is used to display FlowTracker2 handheld software and real-
time graphical display of raw data.
•
Measurement Location
– At each measurement location, the FlowTracker2
records one-second velocity data for the specified averaging time, location, water
depth parameters and a variety of statistical and quality control data.
•
Ping Rate
– The number of pings per second (Hz). The FlowTracker2 ping rate
is 40 Hz.
•
Ping
– A single estimate of the 2D or 3D water velocity.
•
Quality Control Data
– In addition to velocity, the FlowTracker2 records several
quality control parameters. These include signal-to-noise ratio (SNR), standard
error of velocity, boundary adjustment, the number of spikes filtered from data,
and velocity angle. For details about quality control data, see
details.
•
Salinity
– Water salinity is a user-supplied value that is used for sound speed
calculations. Note: If using the system in salt water, a zinc anode should be
installed on the probe for corrosion protection described in
•
Sample
– A sample refers to the mean of 20 pings to produce a measurement of
the 2D or 3D water velocity. A sample includes velocity and signal to noise ratio
data. The FlowTracker2 records two samples per second.
•
Signal Strength
– This refers to the strength of the reflected acoustic signal. It is
a function of the acoustic conditions of the water – primarily the amount and type
of suspended material (scatterers) present. This is most commonly accessed as
a signal-to-noise ratio (SNR).
•
Signal-to-Noise Ratio (SNR)
– SNR is the ratio of the received acoustic signal
strength to the ambient noise level. It is expressed in logarithmic units (dB), and
is the most important quality control data for the FlowTracker2, see
•
Sound Speed
– Speed of sound in water (in m/s) is used to convert the Doppler
regarding the effect of sound speed
on velocity data.
Summary of Contents for SonTek FlowTracker2
Page 107: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 106 ...
Page 117: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 116 ...
Page 196: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 195 ii Stationary Measurement ...
Page 247: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 246 Configuration Changes ...
Page 283: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 282 ...
Page 300: ...SonTek a Xylem brand FlowTracker2 User s Manual January 2019 299 ...