GLP/TOOLS
OxiTop
®
Control OC 100
46
ba31114e07
08/2014
Switching the
GLP
func-
tion on or off
7.4.5
GLP
- calibration interval (
- Calinterval
)
Set the time period (1 to 36 months) here. When it expires, the instru-
ment registers that the next test resource monitoring of the measuring
heads is due. After the calibrating interval expires, measuring is
blocked until the calibration is performed or the
Off
setting is selected.
SETTINGS
menu:
Use
<
><
>
to select the
-
GLP
menu item.
GLP
menu:
Preselected:
- GLP
Confirm with
<RUN/ENTER>
.
Use
<
><
>
to select either
GLP
On
or
Off
.
Confirm with
<RUN/ENTER>
.
RUN/ENTER
If the end of the calibrating interval set up lies within the measuring
time of a measurement to be started, this measurement cannot be
started.
GLP
menu:
Use
<
><
>
to select the
-
Calinterval
menu item.
Confirm with
<RUN/ENTER>
.
Use
<
><
>
to set the time
period for the calibrating interval
(1 - 36 months).
Default setting:
12 months.
Confirm with
<RUN/ENTER>
.
RUN/ENTER