11.4 Cond measurement settings
11.4.1 Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
<MENU/ENTER>
key. After completing the
settings, switch to the measured value display with
<M>
.
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (
TetraCon 925
,
LR 925/01
).
Salinity
0.0
... 70.0
Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if the
automatic salt content correction is
switched off and the manual salt con-
tent correction is switched on.
Response time t90
30
... 300
Response time of the signal filter (in
seconds).
A signal filter in the sensor reduces
the limits of variation of the measured
value. The signal filter is character-
ized by the response time t90. This is
the time after which 90 % of a signal
change is displayed.
This menu item is only available if
sensor and meter support this func-
tion.
You can update the firmware of sen-
sor and meter (see chapter 17).
Reset
-
Resets all sensor settings to the deliv-
ery condition (see section 11.7.1).
Menu item
Possible
setting
Explanation