What to do if...
Multi 3620 IDS
100
ba77169d06
10/2021
Data transmission to
USB printer does not
work
Error message,
Only use
rechargeable
batteries!
Ni-MH 1.2 V, >2000
mAh
Error message,
Memory error 1
15.2
pH
Error message
OFL, UFL
Cause
Remedy
– The USB-B interface is con-
nected to a PC
– Disconnect the PC from the USB-B
interface
– Connected USB printer does not
– Use a suitable USB printer (see sec-
– Check the printer settings (see sec-
Cause
Remedy
– A battery voltage untypical for
NiMH batteries was identified.
– Disconnect the power pack from the
meter.
– Check whether the batteries in the
battery compartment are suitable
(
Ni-MH 1.2 V, >2000 mAh
).
–
Unsuitable batteries
:
Insert suitable batteries and recon-
nect the power pack.
–
Suitable batteries
:
Connect the power pack and close
the error message with
<ESC>
or
<M>
.
CAUTION
The charging of unsuitable battery types may cause damage.
Cause
Remedy
– Instrument storage was not
found
–
Please contact the service.
Cause
Remedy
IDS-pH sensor:
– Measured value outside the
measuring range
– Use suitable IDS-pH sensor
– Air bubble in front of the junction
– Remove air bubble
– Air in the junction
– Extract air or moisten junction