August 2020 – TD 316 OPERATING MANUAL – Motus Wave Sensor 5729
Page 51
3.10.1
Mandatory
Figure 3-25: Mandatory in User Maintenance
All sensors are given a
Node Description
text like
Motus #xxx
(where xxx is the serial number of the sensor). The user
can modify this node description text if required. Be aware that the node description changes to
*Corrupt
Configuration
if it has lost the configuration in flash. Contact the factory if this happens. The configuration is saved in
two sectors in flash memory. A flash sector can be corrupted if the power is lost during the saving of new configuration.
The double flash sector saving ensures that it does not lose the configuration. If one of the sectors is corrupted, the
other sector is used and also saved to the corrupt sector.
3.10.2
Site Info
Figure 3-26; Site Info in User Maintenance
Site Info
is optional information to be entered to store information about the
Owner
. This setting is not used in
calculation.
3.10.3
Serial Port
Figure 3-27: Serial Port settings in User Maintenance
The
Serial Port
group contains setting that deals with the serial port setting
. Interface
is only available as RS-232.