UM Temp74 PT100 Vio EN rev.1 19.05.2023 

 

Offset adjustment  

All instruments are factory pre-calibrated and remain stable over time, but in the event of a reading error, the 

measuring chain can be recalibrated to eliminate this error. 

In the setup menu - Temp Settings - Probe Offset it is possible to adjust the measurement of each individual 

channel independently or to align the measurements of all channels to the reference channel. 

Example: Single channel calibration at 0.0 °C 

To recalibrate the instrument at 0.0°C, proceed as follows: 

 

prepare ice with double-distilled water; 

 

put ice cubes in a 500 ml beaker up to half the volume; 

 

add double-distilled water to ¾ of the volume; 

 

place the beaker on a magnetic stirrer under gentle agitation; 

 

immerse the probe in the beaker, ensuring that it is immersed in water and does not touch the beaker wall; 

 

cover the beaker with polystyrene to isolate the water/ice mixture from the air as much as possible; 

 

leave to stir for about 10 minutes; 

 

wait until the temperature reaches 0°C. 

If the instrument reads 0.0°C (± measuring chain accuracy), this means that the measuring chain is correctly 

calibrated. Otherwise: 

 

access the Setup menu

 Temp Settings

 Offset probes; 

 

select the channel to be adjusted; 

 

the display will show the current temperature, which is to be corrected, and the adjustment value, with 

flashing cursor; 

 

using the keys

 correct the value to 'Current Temperature' to 0.0 °C; 

 

press

 to confirm and record the adjustment. 

Note

: The adjustment affects the entire measurement scale, e.g. if you correct by -1.0°C, then the 

measurement shifts by -1.0°C over the entire range.

 

The setting made in this way remains in memory 

both after switching off the instrument and in case of

 

replacement of batteries.

 

If the sensor is replaced, we recommend resetting the factory calibration from the Temp Settings 

menu

 Reset and proceed with a new calibration if necessary.

 

If the sensor is replaced, we recommend resetting 

the factory calibration from the Temp Settings menu

 Reset and proceed with a new calibration if necessary.

 

If there is no probe connected to the selected channel, the offset cannot be changed. 

 

 

D Software 

 

It is possible to connect the Temp 74 to a PC and use the D software to perform data downloads, 

recording directly on the PC and exports to Excel and PDF. 

The software can be downloaded free of charge from the site: 

 

https://www.giorgiobormac.com/it/download-software_Download.htm. 

 

Connect the USB cable inside each package to the USB port on the top panel of the instrument and the other 

end to a COM port on your computer. 

 

Only use the USB cable supplied with each instrument. 

 

Start the programme and then switch on the instrument. 

 

Wait for the connection to be established (connection data are shown at the bottom left of the display). 

 

 

 

 

 

 

 

 

 

Summary of Contents for Temp74 PT100 Vio

Page 1: ...INSTRUCTION MANUAL Temp74 PT100...

Page 2: ...tilingual copy Espa ol Para reducir los gastos este manual ha sido impreso exclusivamente en italiano e ingl s Contacta tu distribuidor para recibir la copia digital multi idioma Deutsch Um umweltfreu...

Page 3: ...nstrumental Characteristics 6 Technical Data 6 Description Instrument 6 Display single channel 6 Multiview screen 7 Keyboard 7 Key Functions 8 LED 8 Installation 8 Supplied Components 8 Commissioning...

Page 4: ...ngs 14 Date and time 14 Units of Measurement 14 Language 14 Auto off 14 Sound 14 Password 15 Temperature measurement 15 Measure 15 MAX MIN function 15 HOLD function 15 Other functions 15 Automatic swi...

Page 5: ...ormation NOTE for useful product information Warning symbols Additional documents providing security information The following documents can provide the operator with additional information to work sa...

Page 6: ...mended to occasionally clean the instrument casing with a damp cloth and mild detergent This operation should be carried out with the instrument switched off and only by experienced and authorised per...

Page 7: ...MAX MIN reading Yes Offset Adjustment Yes for each channel independently Internal memory 200 000 data Data logger Manual Automatic Programmed Sampling 1 min 99 hours Working conditions Temperature 0 6...

Page 8: ...X and MIN values recorded Press for 3 seconds to activate and deactivate the HOLD function and to deactivate the MAX MIN recording function if active Directional key DOWN RM to scroll through menus an...

Page 9: ...he display brightness Only at this point do the buttons regain their function LED All instruments are equipped with a two colour LED red and green that provides the user with important information on...

Page 10: ...carrying case Only use the original carrying case to transport the instrument Contact your local distributor if you need to buy it back The inside of the case is shaped to accommodate the instrument...

Page 11: ...e below shows the Setup menu structure for the parameter for each programme the options the user can choose and the default value Description Options Factory Settings Sensor ID Probe 1 Probe 2 Probe 3...

Page 12: ...annels to be aligned to the reference channel Select our reference channel from the Setup menu Temp Settings Reference Probe and press to confirm In the Offset Probes menu select Align All The measure...

Page 13: ...the table above Note When the reset is complete the instrument returns to measurement mode and the parameters are reset to default settings Log Settings Menu TEMP Measurement Settings LOG Data logger...

Page 14: ...pendently once the set date and time is reached and will stop Note If recording takes place while there is a single channel measurement on the display data concerning only that channel will be saved I...

Page 15: ...er battery life Brightness Access this Setup menu to choose between three different display brightness levels Low Normal High Note Keeping the display at high brightness at all times negatively affect...

Page 16: ...appears on the display Always wait for the measurement to become stable Note If no temperature probe is connected the message CONNECT A PROBE appears When connecting or disconnecting probes the instru...

Page 17: ...corrected and the adjustment value with flashing cursor using the keys correct the value to Current Temperature to 0 0 C press to confirm and record the adjustment Note The adjustment affects the ent...

Page 18: ...e manufacturer will repair or replace defective components This guarantee applies only to the electronic part and does not apply if the product has been damaged misused exposed to radiation or corrosi...

Page 19: ...18 UM Temp74 PT100 Vio EN rev 1 19 05 2023 XS Instruments Via della Meccanica n 25 41012 Carpi MO ITALY Tel 39059 653274 Fax...

Reviews: