15   

UM Temp74 PT100 Vio EN rev.1 19.05.2023 

 

Password 

Through this menu it is possible to enter a 4-digit password, which will be required later to enter the Temp and 

Log menus, in order to safeguard sensitive settings. To enter or change the password, proceed as follows: 

 

access the Password menu; 

 

enter the current password (*) using

 to change the digit with the blinking cursor, and press

 to confirm 

and move to the next digit; 

 

the words change password appear: you can now enter the desired password; 

 

press the

 button until you return to the measurement screen. 

The password is now set correctly and will be required to access the TEMP settings and LOG settings menus. 

Note: 

*When first set up after purchase, the factory password is 0000. 

To deactivate the password lock, set password: 0000.

 

 

 

Temperature measurement 

 

 

Measure 

 

Switch the instrument on with the

 button. 

 

When switched on, they are displayed in order: 

- start page with software version; 

- calibration data page: the calibration data of each channel are shown; 

- if several probes are connected, the instrument shows the multiview screen; 

- if only one probe is connected, the dedicated channel page is accessed. 

 

Place the probe(s) at the point to be measured and wait for stability. 

 

When the reading becomes stable, the icon

 appears on the display.  Always wait for the measurement to 

become stable. 

Note 

If no temperature probe is connected, the message 'CONNECT A PROBE' appears

When connecting or disconnecting probes, the instrument will automatically update the measurement 

screen to show the temperatures of the connected probes on the display.   

 

 

MAX/MIN function 

During measurement, press the

 button to display the minimum and maximum values. 

The following symbols will appear on the display:  

 

 

The values are updated in real time, as long as the function remains active. 

To deactivate the MAX/MIN function, press the

 button, the values will disappear from the display. 

Note

: If auto switch-off is set, then the instrument will switch off after 20 minutes. If you intend to record minimum 

and maximum values over a longer period of time, it is necessary to disable the auto switch-off. 

MAX/MIN values are not saved in the instrument's memory and cannot be exported.  

However, they can be easily obtained from the Excel file available on the D. 

 

 

HOLD function 

During measurement, press and hold the

 button for 3 seconds to freeze the reading. HOLD appears on the 

display and the reading is no longer updated. Press and hold for 3 seconds to unlock the reading. 

 

 

Other functions 

 

 

Automatic switch-off 

The instrument switches off automatically after 20 minutes of inactivity. 

To disable auto power off, go to Setup

 Settings

 Auto Power Off and select No. 

 

Maximum measured value 

Minimum measured value 

Summary of Contents for Temp74 PT100 Vio

Page 1: ...INSTRUCTION MANUAL Temp74 PT100...

Page 2: ...tilingual copy Espa ol Para reducir los gastos este manual ha sido impreso exclusivamente en italiano e ingl s Contacta tu distribuidor para recibir la copia digital multi idioma Deutsch Um umweltfreu...

Page 3: ...nstrumental Characteristics 6 Technical Data 6 Description Instrument 6 Display single channel 6 Multiview screen 7 Keyboard 7 Key Functions 8 LED 8 Installation 8 Supplied Components 8 Commissioning...

Page 4: ...ngs 14 Date and time 14 Units of Measurement 14 Language 14 Auto off 14 Sound 14 Password 15 Temperature measurement 15 Measure 15 MAX MIN function 15 HOLD function 15 Other functions 15 Automatic swi...

Page 5: ...ormation NOTE for useful product information Warning symbols Additional documents providing security information The following documents can provide the operator with additional information to work sa...

Page 6: ...mended to occasionally clean the instrument casing with a damp cloth and mild detergent This operation should be carried out with the instrument switched off and only by experienced and authorised per...

Page 7: ...MAX MIN reading Yes Offset Adjustment Yes for each channel independently Internal memory 200 000 data Data logger Manual Automatic Programmed Sampling 1 min 99 hours Working conditions Temperature 0 6...

Page 8: ...X and MIN values recorded Press for 3 seconds to activate and deactivate the HOLD function and to deactivate the MAX MIN recording function if active Directional key DOWN RM to scroll through menus an...

Page 9: ...he display brightness Only at this point do the buttons regain their function LED All instruments are equipped with a two colour LED red and green that provides the user with important information on...

Page 10: ...carrying case Only use the original carrying case to transport the instrument Contact your local distributor if you need to buy it back The inside of the case is shaped to accommodate the instrument...

Page 11: ...e below shows the Setup menu structure for the parameter for each programme the options the user can choose and the default value Description Options Factory Settings Sensor ID Probe 1 Probe 2 Probe 3...

Page 12: ...annels to be aligned to the reference channel Select our reference channel from the Setup menu Temp Settings Reference Probe and press to confirm In the Offset Probes menu select Align All The measure...

Page 13: ...the table above Note When the reset is complete the instrument returns to measurement mode and the parameters are reset to default settings Log Settings Menu TEMP Measurement Settings LOG Data logger...

Page 14: ...pendently once the set date and time is reached and will stop Note If recording takes place while there is a single channel measurement on the display data concerning only that channel will be saved I...

Page 15: ...er battery life Brightness Access this Setup menu to choose between three different display brightness levels Low Normal High Note Keeping the display at high brightness at all times negatively affect...

Page 16: ...appears on the display Always wait for the measurement to become stable Note If no temperature probe is connected the message CONNECT A PROBE appears When connecting or disconnecting probes the instru...

Page 17: ...corrected and the adjustment value with flashing cursor using the keys correct the value to Current Temperature to 0 0 C press to confirm and record the adjustment Note The adjustment affects the ent...

Page 18: ...e manufacturer will repair or replace defective components This guarantee applies only to the electronic part and does not apply if the product has been damaged misused exposed to radiation or corrosi...

Page 19: ...18 UM Temp74 PT100 Vio EN rev 1 19 05 2023 XS Instruments Via della Meccanica n 25 41012 Carpi MO ITALY Tel 39059 653274 Fax...

Reviews: