background image

15

Copyright ©2011 XS Scuba

 

Service & Repair Manual

a.  If airflow cannot be heard, it will be necessary to initi-

ate a slight leak between the orifice and poppet.  Press 
the knurled knob inward to engage the slotted drive of 
the Inline Adjustment Tool with the head of the orifice 
inside the inlet fitting.  Then, turn the orifice slightly 
counter-clockwise while lightly depressing the purge 
cover to prevent wear on the seat.  Do not adjust any 
further than is needed to establish a slight leak.

NOTE:

  If more than a slight adjustment is required to 

initiate a leak, it is important to check the adjustment of 
the locknut on the poppet shaft to verify once again that 
it is set correctly to its preliminary setting, as specified 
in Step 8 of the Reassembly Procedure.

5.  Hold the drive of the Inline Adjustment Tool engaged with 

the orifice.  While lightly depressing the purge cover, turn 
the orifice slightly clockwise a very small fraction of a turn.  
Pause after each adjustment to listen, and be careful to 
avoid over-adjusting beyond the point that the leak has 
stopped.  When the leak has stopped, purge the second 
stage several times to ensure that it does not return.  Ob-
serve the IP Test Gauge while purging the second stage to 
verify that it indicates a stable intermediate pressure after 
each cycle, with no creep or fluctuation.

CAUTION

:    Over-adjustment  of  the  orifice  can  cause 

excessive spring load in the second stage valve, and 
may impair the regulator’s performance.

6.  While holding the second stage with the mouthpiece fac-

ing down, gently shake it up and down.  Listen closely 
to determine whether any movement of the lever can be 
heard inside, indicating that the lever has dropped.  If lever 
movement cannot be detected, proceed to step 8.  Other-
wise, perform the following steps to raise the lever:
a.  Remove the retaining ring, purge cover, thrust washer 

and diaphragm to expose the valve assembly.

b.  Apply a 4" open-end wrench to turn the locknut clock-

wise a small fraction of a turn.  Do not over-adjust, 
since it is important to tighten the nut onto the poppet 
only as far as necessary to remove lever slack.

c.  Repeat the above step until lever slack is eliminated.
d.  Reinstall the diaphragm, thrust washer, purge cover, 

and retaining ring.

Fig. 13 – Inline Adjustment Tool w/ IP Gauge

Summary of Contents for AltAir Octo

Page 1: ...Service Repair Manual SERVICE AND REPAIR MANUAL ALTAIR OCTO PN RG300 Revised 02 11...

Page 2: ...Tools 5 Section 2 Preliminary Inspection External Inspection 6 Immersion Leak Test 6 Section 3 Disassembly Procedures General Guidelines 7 Section 4 Reassembly Procedures General Guidelines 11 Sectio...

Page 3: ...product it is extremely critical that it receives service according to the procedures outlined in this manual on a regularly scheduled basis at least once a year with normal or infrequent use NOTE A u...

Page 4: ...lt from improper use of this tool Be sure to follow the instructions pro vided in this manual NOTE XS Scuba especially endorses and recommends the Jaws Gripmaster Bench Vise available from Peter Built...

Page 5: ...hat are designed and manufactured specifically for servicing SCUBA equipment we recommend one source for all your service facility s needs Peter Built Co in Galion Ohio is dedicated to serving the div...

Page 6: ...3 Check the overall condition of the mouthpiece exhaust grill and other external parts to evaluate the type of use and maintenance the regulator has received Immersion Leak Test 1 Check to ensure tha...

Page 7: ...amic o rings After passing close inspection static O rings may sometimes be reused although this is not necessarily recommended Dynamic O rings must be automatically discarded and replaced with every...

Page 8: ...re the replacement of the LP hose 4 Snip the plastic tie strap 6 that holds the mouthpiece 7 and gently pull the mouthpiece off the second stage housing 5 Inspect the mouthpiece to ensure that it is s...

Page 9: ...aining ring is difficult to loosen XS Scuba recommends using the Rim Clamp P N 20 795 200 available from Peter Built Co Complete instructions are provided with the tool 12 Lift off the diaphragm purge...

Page 10: ...into both slots on either side of the second stage housing See Fig 6 Gently pull the grill out of the housing being careful to avoid bending 20 Remove and discard both of the exhaust valves 8 by pulli...

Page 11: ...ends Christo Lube MCG 11 PN MS150 to be used exclusively for the lubrication and dressing of O rings and other internal parts Christo Lube provides superior lubrication to that of silicone grease espe...

Page 12: ...and place the spring over the threaded shaft of the poppet 5 Mate the squared poppet shaft through the lever sup port 17 and then fit the inlet fitting 22 over the poppet head Screw the inlet fitting...

Page 13: ...ce further beyond the point where it makes contact with the low pressure seat Doing so may wear the LP seat or sealing surface of the orifice and can also result in an incorrect adjustment of the seco...

Page 14: ...se 26 Mate the female fitting of the LP hose onto the male fitting of the IP Test Gauge and tighten the swivel nut clockwise until finger snug 3 Install the O ring 28 onto the male fitting of the LP h...

Page 15: ...that the leak has stopped When the leak has stopped purge the second stage several times to ensure that it does not return Ob serve the IP Test Gauge while purging the second stage to verify that it...

Page 16: ...e fitting to 55 5 inch lbs Subjective Breathing Test 1 Connect the regulator to a cylinder containing 2 500 3 000 psi and open the valve to pressurize the system 2 Fully depress the second stage purge...

Page 17: ...the problem s NOTE The location of extremely small leaks can best be detected by applying a soap solution to the leak area Before disassembling to correct any leaks rinse the entire regulator thoroug...

Page 18: ...ate pressure 1 Refer to first stage troubleshooting 2 Orifice incorrectly adjusted 2 Reset to preliminary setting and readjust 3 Lever is damaged 3 Replace with new 1 Lever is slack or orifice incorre...

Page 19: ...1 00 15 RP300 15 Spacer 16 RP300 16 Lever 17 RP300 17 Lever Support 18 RP300 18 Spring 19 RP300 19 Poppet 20 RP300 20 Seat 21 RP300 21 O ring 22 RP300 22 Inlet Fitting 23 RP300 23 Adjustable Orifice 2...

Reviews: