Instrumental Features

 

 

Datasheet                                                                                                   

 

 

 

Instruments

 

 

 

Temp7 PT100 Vio

 

Temp7 K/T Vio

 

Temp7 NTC Vio

 

Temp

er

at

u

re

 

Probes

 

PT100

 

Thermocouple K e T

 

NTC 30KΩ

 

Input

 

Single connector 3 pin

 

Single connector ANSI

 

Single connector 

CINCH / RCA

 

Range of 
measurement

 

-200…+999 °C

 

Thermocouple K: 

 

-200…+1350 °C

 

Thermocouple T: 

 

-250…+400 °C

 

-50,0…+150,0 °C

 

Resolution

 

0,1 °C 

 

(da -99,9 a +199,9)

 

1 °C on the full scale

 

0,1 °C 

 

(da -99,9 a +199,9)

 

1 °C on the full scale

 

0,1 °C

 

Accuracy

 

±2 °C (-200…-100 °C)

 

±0,2 °C (-99,9…+199,9 °C)

 

±2 °C (+200…+999 °C)

 

±0,1% of the read / ±0,4 °C 

(over –150 °C)

 

±0,25% of the read / ±1 °C 

(below –150 °C)

 

±0,2 °C

 

Sy

st

em

 

Display

 

High-definition colours LCD, Backlit

 

Brightness setup

 

Manual

 

Sleep Mode

 

Yes

 

Auto turn off

 

Yes, 20 minute (deactivable)

 

LED Indication

 

Yes

 

Measurement stability

 

Yes

 

HOLD function

 

Yes

 

Read MAX/MIN

 

Yes

 

Offset regulation

 

Yes

 

Working conditions

 

Temperature: 0…+60 °C

 

Humidity: <95% U.R. (not steamed)

 

Maximum altitude

 

2000 m

 

Dimensions

 

185 x 85 x 45 mm

 

Weight

 

400 g

 

IP grade

 

IP 56

 

Sound level during 
standard operation

 

< 80 dB

 

Power supply

 

3 x 1,5V “AA” alkaline

 

Battery life

 

> 550 hours

 

Reference legislation

 

EMC 2014/30/UE

 

RoHS 2011/65/EU

 

EN 61326-1

 

 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for 7 Vio

Page 1: ...USER MANUAL Temp7 PT100 Temp7 K T Temp7 NTC...

Page 2: ...1 UM Termometri serie 7Vio ENG rev 0 14 04 2021...

Page 3: ...Features 6 Datasheet 6 Instrument Description 7 Display 7 Keypad 7 LED 8 Setup 8 Supplied component 8 Start up 8 Switching On Off 8 Batteries replacement 8 Instrument transportation 9 Keys Function 9...

Page 4: ...serie 7Vio ENG rev 0 14 04 2021 Warranty period and limitations 16 Disposal of electrical devices 16 XS Instruments Via della Meccanica n 25 41012 Carpi MO ITALY Tel 39059 653274 Fax 39059653282 www x...

Page 5: ...ON for a medium risk hazardous situation which could lead to serious injury or death if not avoided ATTENTION for a dangerous situation with reduced risk which can cause material damage data loss or m...

Page 6: ...ime in adverse conditions exposure to direct light heat sources or places saturated by gas or vapours or in environments with conditions different from those mentioned in this manual Device maintenanc...

Page 7: ...curacy 2 C 200 100 C 0 2 C 99 9 199 9 C 2 C 200 999 C 0 1 of the read 0 4 C over 150 C 0 25 of the read 1 C below 150 C 0 2 C System Display High definition colours LCD Backlit Brightness setup Manual...

Page 8: ...uipment ESC key to go back at the measurement mode OK Setup key to confirm the chosen values and enter the Setup from the measurement mode DOWN HOLD key to scroll menus and values in the Setup Short p...

Page 9: ...witching On Off Turn on the system by pressing the button The display initially activates all segments and then appears model and firmware of the instrument settings related to the most important para...

Page 10: ...go back to measurement mode Short MAX MIN memory activation MAX value display and MIN value display It is used to scroll and change the parameters of the Setup menu Long Press 3 sec 3 seconds pressure...

Page 11: ...s and icons Symbols Description Type of probe connected Error in measurement Battery charge indication Measurement stability indicator The bars scroll if the measurement is not stable Press the direct...

Page 12: ...SETUP mode select the parameter you want to edit by using the directional keys and confirming with Within the selected menu move between the different programs using the directional buttons and press...

Page 13: ...e temperature probe connected If menu P 7 2 is accessed without a temperature sensor connected the display will show the error message CONNECT THE PROBE accompanied by the flashing icon Note For the T...

Page 14: ...n three different levels of display brightness LOW low NORMAL medium HIGH high Note Keeping the display bright always adversely affects battery life P9 5 Sleep mode Access this setup menu to select wh...

Page 15: ...ill flash MIN MAX REC Press the key once to see the maximum value reached up to that moment press it another time to see the minimum value reached up to that moment Pressing again will return the inst...

Page 16: ...isolate the water ice mixture from the air as much as possible stir for about 10 minutes in this way a temperature of 0 C is reached If the instrument reads 0 0 C measurement chain accuracy then the i...

Page 17: ...urer will repair or replace defective components This warranty is valid only and exclusively on the electronic parts of the device and does not apply if the product has been damaged used incorrectly e...

Reviews: