XONOX CT 200 Quality Assurance Document Download Page 1

 

 
 
 

Quality assurance document for the CT 200 measuring 

instrument 

 

If you have any questions do not hesitate to contact us at any time. 

Mail:  [email protected] // Phone:  +49 6441 21049-13 

 
 
 

Check of mechanical alignment

 

1)  Place 10mm gauge block on the lens transducer and tighten 

the chuck. 

 
 
 

 

2)  Switch on the digital dial gauge and hold the SET / ZERO button for two seconds. Now 

in the display a10.0000 mm and P flashes. Pressing the Set button to set this 
value (+10.0000 mm). 

 

 

 

3)  Operate the left switch (the measuring tips touch the 10mm gauge block). It is important 

that the upper measuring tip first touch the gauge block. Then, press the SET button 
on the remote controller. The gauge is now calibrated for +10.0000 mm. 

 

 

 

 

 

Summary of Contents for CT 200

Page 1: ...ransducer and tighten the chuck 2 Switch on the digital dial gauge and hold the SET ZERO button for two seconds Now in the display appears 10 0000 mm and P flashes Pressing the Set button to set this value 10 0000 mm 3 Operate the left switch the measuring tips touch the 10mm gauge block It is important that the upper measuring tip first touch the gauge block Then press the SET button on the remot...

Page 2: ... mm and make some measurements again The value must be 10 0000 mm 6 To know how save your repeatability measurement is you can form the arithmetic mean value with this formula 𝑥 𝑥1 𝑥2 𝑥𝑛 n 7 If you have big deviations please check the situation quality of the special measuring tips It is important that you use only the Xonox measuring tips no ruby or similar 8 To control the mechanical axis demoun...

Page 3: ...Clamp the measuring sleeve in the chuck and let the upper and lower shaft glide into the measuring sleeve For this action the upper boom can be held by hand and the lower boom can be operated with the pneumatic left switch When you see that the lower shaft entering into the measuring sleeve lead the upper shaft by hand in the measuring sleeve 12 When the alignment is centered both shafts slide int...

Page 4: ...iBall Ø 25 400 Tolerance for the measurements are 4 µm 14 It is absolutely forbidden to set the dial gauge to zero when the measuring tips stay in contact In additional it is very important that the alignment of the CT don t be checked in this way 15 Use always the gauge block to set the digital dial gauge to zero ...