XO 4 & XO FLEX TROUBLE SHOOTING GUIDE
Version 3.30
78
XO FOOT CONTROL
Failure description
Reason
Solution
Display shows « Foot
control Error” during start-
up
There is no BUS RS-485
communication between the foot
control and the Unit
Verify that the foot control cable
(CA-009) is not damaged.
Measure with an ohmmeter from
the Backplane J14 to the foot
control J1 pin 1 to 4, if there a
break on the cable CA-009 Value
MΩ.
Measure the resistance between
Pin 1 and 2 (RS-485
communication BUS, the value
should be around 120Ω.
Measure the 5V voltage between
Pin 1 () et Pin 4 (Green
GND).
If the +5V Voltage is missing :
-
Verify the fuses on “Power
Supply AN-371”
-
Read “5V Overvoltage
protection” chapter “Power
Supply AN-371” on page 14
If all the above is correct, then the
foot control is defective.
All defective foot control must be
sent back to XO-care
“Network Fail!! “Call
Service”
Displayed after switching on the
unit if there is no communication
between the "Stand Controller"
A0-981 and the pedal.
Either due to a cable break, a
Recalibrate the foot control.
Check the CA-003
communication cable is properly
connected between J14 (bridge)
and J32 (Backplane) in the unit.
Summary of Contents for 4
Page 1: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE YB 907...
Page 30: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 30 TECHNICIAN MENU Configuration Adjustments...
Page 31: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 31 SUSPENSION CONFIGURATION...
Page 54: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 54...
Page 58: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 58 LIQUID SENSOR PCB AP 869...
Page 64: ...XO 4 XO FLEX TROUBLE SHOOTING GUIDE Version 3 30 64 SUCTION DISINFECTION SYSTEM...