Alarms and Errors
M370078-01
3-19
3
Clearing Alarms
Clearing Triggered and Manual Alarms
To clear a triggered alarm, use one of the following methods:
• Turn the power supply Off and then On.
• Press and hold the rotary Adjust/Enter control for 3 seconds.
• Press the reset key combination (Flash Fail ONLY). See “Resetting the
Power Supply” on page 3–39.
To clear a manual alarm:
1. Press and hold the rotary Adjust/Enter control for 3 seconds.
The
ALr CLr
message will be displayed on the front panel for a few
seconds.
2. Repeat the alarm clearing until you have cleared all alarms.
Clearing a Flash Failure Alarm
The Flash Failure alarm is the only exception for clearing alarms. Flash
failure can only be cleared by a reset (see “Resetting the Power Supply”
on page 3–39) or by power cycling the unit.
To clear a triggered alarm, use one of the following methods:
•
Turn the power supply Off and then On.
•
Press and hold the rotary Adjust/Enter control for 3 seconds.
•
Press the reset key combination (Flash Fail ONLY). See “Resetting
the Power Supply” on page 3–39.
The SCPI command (s) for these instructions are:
[:]OUTPut[<channel>]:PROTection:CLEar
Important:
When an alarm is cleared by using this method, only the current
highest precedence alarm is cleared. If another alarm has occurred, then this
alarm will now become the highest precedence alarm and will begin blinking on
the display.
Summary of Contents for Watt Series Programmable DC Power Supply XG...
Page 2: ......
Page 3: ...XG 850 Watt Series Programmable DC Power Supply Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 30: ...1 8...
Page 52: ...Local Operation 3 8 M370078 01 Figure 3 3 Front Panel Menu System...
Page 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370078 01...
Page 198: ...5 72...
Page 262: ...A 36...
Page 284: ...IX 4...
Page 285: ......