 mini+: Quick Start Guide

 

 

 

 

SIDE 1 

 

Warning 

– Refer to Safety Precautions Booklet 035-00636R for all safety precautions prior to use.   

User Guide 035-00587R download

https://www.excelitas.com

 

Applies to X-Cite mini Series models:  XMPS, XMPL, XTMS, XTML, XT120Lm 

 

 

WARNING: Never look at operating LED!  The light could severely damage the cornea and retina of the eye if the light is observed directly. Eye shielding 
must be used at all times as well as clothing to protect exposed skin  

Printed in Canada 

Refer to user guide for more details. 

035-00692R Rev. 2 

 

SETUP 
Unpack 

1.

 

Carefully unpack the unit and accessories from shipping carton.   

2.

 

When removing the LED Head and miniCUBE from the carton, ensure that 

both components are supported and there is always some slack in the 

cable.  Do not use the cable as a “handle”.   

 

Install the Microscope Flange 

1.

 

Remove protective cap from LED Head.   

2.

 

Align the mounting holes and dowel pins on the Microscope Flange and the 
LED Head. Do not touch the lens in the Microscope Flange.   

3.

 

Attach the Microscope Flange using the Hex Tool and three (3) fasteners 

provided.   

 

Install on Microscope 

1.

 

Position the equipment.  Do not kink or bend cable.  Ensure adequate 

clearance for ventilation around miniCUBE. 

2.

 

Connect LED Head to microscope light train. Verify that the LED Head is 

oriented with the arrow label (rear side of the LED Head) is pointing up.  

3.

 

Connect speedDIAL cable to “remote” port on miniCUBE.   

4.

 

Connect AC power to properly grounded outlet.   

 

MANUAL OPERATION 

1.

 

Flip rocker switch to turn unit on.   

2.

 

When display shows “x%”, it is ready to use.  

3.

 

Click speedDIAL to illuminate a specimen. 

4.

 

Turn speedDIAL to adjust intensity. 

 
REGULATORY

 (refer to user guide for complete details) 

 
Product Safety and Electromagnetic Compatibility 

 

The X-Cite mini+ has been tested and found to comply with product safety and 

electromagnetic compatibility requirements.  For a complete list of tests and for 

certification details, please contact your X-Cite representative or visit 

https://www.excelitas.com

 

 

CE Marking 

 

 

Council Directive 2014/35/EU 

  Low Voltage Directive 

 

Council Directive 2014/30/EU   

  EMC Directive 

 

Council Directive 2012/19/EU 

  WEEE Directive 

 

Council Directive 2011/65/EU 

  RoHS 

as amended by (EU) 2015/863 

 

 

 

BASIC TROUBLESHOOTING

 (refer to user guide for complete details)  

 

Category 

Symptom 

Action 

Error 

messages 

Error 1 (High Temp.) 

Turn off, let cool; check 

clearance around vents 

Error 3 or 5 (Internal) 

Turn off, restart   

Error 6 (Low Temp.) 

Allow unit to warm to room 
temperature and restart 

Intensity 

Low output intensity 

Check speedDIAL settings 

Check flange type 

Check microscope beam path, 

filters, apertures, shutters, 

objectives  

speedDIAL 

Double-click turns off LED  

Increase delay in the speed 

(Spd) setting 

New settings forgotten when 

unit is shut down  

After changing settings, wait at 

least five minutes before 

shutting unit down 

Dial does not scroll through 

the menu  

Turn dial in the opposite 

direction 

“Lock” icon on display, 

speedDIAL not responding to 

manual control 

Send “unlock” command via 

computer OR power down unit 

and restart 

TTL 

TTL does not trigger unit 

Enable TTL mode 

TTL response time is longer 

than usual 

Verify that timeout settings are 

appropriately set   

UV 
Excitation 

No DAPI signal is visible, all 
other fluorophores are fine   

Unit is in “UV Off” mode   
Enable “UV On” mode   

        

 

        

                              

                     

                   

 