background image

                                                                                                                                                            

SIDE 1

 

X-Cite

®

 120LED: Quick Start Guide

 

 

Warning – Refer to User Guide 035-00536R for all safety precautions prior to use.

 

 

 

WARNING:

 Possibly hazardous optical radiation emitted from this product.  Avoid eye and skin exposure to unshielded product.  Do not 

look at operating LED.  Eye injury may result.  IEC 62471:2006 / EN62471:2008 Risk Group 3.                                                  

Refer to X-Cite 120LED User’s Guide 035-00536R, Section 2

 

Printed in Canada 
035-00537R Rev. 0 

 

SETUP 
Unpack 

1.

 

Carefully unpack the unit and accessories from shipping carton.   

2.

 

When removing the LED Head and powerCUBE from the carton, 
ensure that both components are supported and there is always 
some slack in the cable.  

Do not use the cable as a “handle”.

   

 

Install the Microscope Flange 

1.

 

Remove protective cap from LED Head.   

2.

 

Align the mounting holes and dowel pins on the Microscope 
Flange and the LED Head.   

3.

 

Attach the Microscope Flange using the Hex Tool and three (3) 
fasteners provided.   

 

Install on Microscope 

1.

 

Position the equipment.  Do not kink or bend cable.  Ensure 
adequate clearance for ventilation around powerCUBE. 

2.

 

Connect LED Head to microscope light train.  Verify “arrow” is 
pointing up, and vents are on bottom of LED Head. 

3.

 

Connect speedDIAL cable to “remote” port on powerCUBE.   

4.

 

Connect AC power to properly grounded outlet.   

 

MANUAL OPERATION 

1.

 

Flip rocker switch to turn unit on.   

2.

 

When display shows “x%”, it is ready to use.  

3.

 

Click speedDIAL to illuminate a specimen. 

4.

 

Turn speedDIAL to adjust intensity.  

 

BASIC TROUBLESHOOTING  

(refer to user guide for complete details)

  

 

Category 

Symptom 

Action 

Error 
messages 

Error 1 (High Temp.) 

Turn off, let cool; check 
clearance around vents 

Error 3 or 5 (Internal) 

Turn off, restart   

Error 6 (Low Temp.) 

Allow unit to warm to room 
temperature and restart 

Intensity 

Low output intensity 

Check speedDIAL settings 
Check flange type 
Check microscope beam path, 
filters, apertures, shutters, 
objectives  

speedDIAL 

Double-click turns off LED  

Increase delay in the speed 
(Spd) setting 

New settings forgotten 
when unit is shut down  

After changing settings, wait 
at least five minutes before 
shutting unit down 

Dial does not scroll 
through the menu  

Turn dial in the opposite 
direction 

“Lock” icon on display, 
speedDIAL not responding 
to manual control 

Send “unlock” command via 
computer OR power down 
unit and restart 

TTL 

TTL does not trigger unit 

Enable TTL mode 

TTL response time is 
longer than usual 

Verify that timeout settings 
are appropriately set   

 