800.255.3126 | wyeastmedical.com
51925 R06 Aerial Mobile Full Body Lift
Operator’s Manual
Page 9 of 20
AERIAL
Operator’s Manual
Battery / Control Box Combo:
The control box disperses the power from the battery to the
leg, lift arm, and handset functions.
The Battery Cap (P/N:
51658
) snaps into place on top of
the battery.
The Battery (P/N Std:
51655
| Upgrade:
52431
) should be
charged for a full 24 hours prior to the first use.
The Battery should be charged after every use.
The Control Box (P/N Std:
51657
| Upgrade:
52432
)
receives the connections from the leg actuators, lift arm
actuator and the handset. Please see Fig. 1 for connections.
Control Box Display:
(Upgrade only)
Standard Display:
Cycles done - number of fully loaded lifts on actuator
Work done - Amp power
Overloads - Number of overloads attempted
Days between service - Number of days since last service
Icons shown when hand control button active
Battery Information:
Battery State: 100% - 50%
Battery State: 50% - 25%
Battery State: Battery needs charging – less than 25%
Charging Battery:
Mains cable plugged in. Lift will not work in this state.
Short circuit
Needs service:
standard set to 12 months / 8000 cycles.
See Periodic Maintenance on page 13 for reset
instructions. System will still function.
Overload has occurred
Standard
Battery
Cap
Battery
Control
Box
Upgrade
NOTE:
For further
information,
please refer to the
LINAK CBJ-Care
operators manual.