Operation
TresCon UNO
5 - 36
ba43134e08
05/2017
5.3.3
Limit indicator: Basic information
With a limit indicator, a relay switches when a specified limiting value is
exceeded or undercut.
Limit indicators can be used in the following way:
Monitoring a limiting value using a relay:
when a limiting value (upper or lower limiting value) is exceeded or
undercut, a relay switches. The
Open
or
Close
relay actions are pos-
sible.
Monitoring two limiting values using two relays:
If the upper limiting value is exceeded or undercut, a relay switches,
and if the lower limiting value is exceeded or undercut, another relay
switches. The
Open
or
Close
relay actions are possible.
Note
If the simple monitoring function (
Open
,
Close
) with one or two relays
is not sufficient, use proportional regulation (see section 5.3.1).
Monitoring limiting
values using one or two
relays
Fig. 5-15 Switching points for relays with the function of a limit indicator
1
Upper limit value (relay 1) exceeded
2
Selected switching delay t1 for relay 1 expired
Relay 1 switches
3
Hysteresis for upper limiting value (relay 1) undercut
4
Selected switching delay td1 for relay 1 expired
Relay 1 switches back
Time
Measured
value
UL
LL
Hysteresis UL
Hysteresis LL
3
6
5
Relay 1
Relay 2
4
1
2
td2
td1
td1
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......