
Maintenance
TresCon UNO
6 - 6
ba43134e08
05/2017
6.1.3
Testing the interfaces
You can send a test string "Test string RS232 (or RS485)" to the selec-
ted interface.
Sending a string
Example Test string
Fig. 6-1
RS485 test string
1
Measuring mode
Call up the
Service
menu: Simultaneously press the keys,
4
and
5
. The request
Enter PIN
appears
.
2
Input the PIN digit-by-digit with
68
. Change the digit
with
57
and confirm with
3
. The
Service Mode
menu appears
.
3
Select the
RS232
or
RS485
menu item with
68
.
4
Confirm the selection with
3
. The test string is sent to the
selected interface. A confirmation appears on the display.
------------------------------------------------------------
RS485 test string
------------------------------------------------------------
TresCon>
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......