5.8
Valve control
General
If the TresCon UNO is operated without sample filtration (e.g. PurCon),
sediments may collect in the overflow vessel. We recommend to use
the overflow vessel with the TC/ ÜB V controllable valve (accessory).
Thus the overflow vessel is emptied automatically; the sediments are
removed.
Note
How to install the TC/ ÜB V overflow vessel with valve is described in
section 3.6.2 from page 3-24.
Note
We recommed to adjust the opening time and interval to the settings of
AutoCal and AutoClean. Thus, a lack of sample cannot occur during
the opening. The valves always open for 5 seconds.
Setting the opening
behavior
Parameter
Values
Valve
1, 2, 3 (number of the valve con-
nection on the terminal strip)
Time of opening
Time (hh:mm)
Opening interval
1, 2, 3, ... 24 h
1
Measuring mode
Call up the
General Parameters
menu: Simultaneously press
the keys,
4
and
2
.
2
Select the
Valve control
menu item with
68
and confirm
with
3
. The prompt
Presentation of data/Change
appears
.
3
Select the
Change
mode with
68
and confirm with
3
.
4
Input the PIN digit-by-digit with
68
. Change the digit
with
57
and confirm with
3
. The
Valve control
menu
appears on the display.
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......