5.4.3
Recorder function, Record of measured values
With the
Record of measured values
function, the recorder output pro-
vides a current that is independent of the measured value. With the
Recorder
,
Recorder Beginning
,
Recorder End Point
and
di/dt
parame-
ters, the output of measured values is defined.
Settings
Note
The setting of the recorder parameters is described in section
5.4.2 S
5.4.4
Recorder function, Controller
Setting
Selection/Values
Explanation
Recorder
0 - 20 mA
4 - 20 mA
Signal range of the
recorder output
Recorder Beginning
within the measu-
ring range
(depending on the
analyzer module)
Note:
Minimum distance:
10% of the measuring
range
Recorder End Point
di/dt
0.1 mA/s
1 mA/s
5 mA/s
10 mA/s
20 mA/s
Attenuation: Speed of
change of the output cur-
rent in the case of an
erratic change of the
input signal.
Response with stop
any in the range
0 ... 20 mA
In the Stop mode, the
current output provides
the defined current
value. Also in the case of
AutoCal
,
AutoClean
etc.
Hold
The current at the output
remains unchanged.
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......