Maintenance
NOx-N
3 - 4
ba43110e05
01/2011
3.2
Filling the system
This routine fills the wet chemical system with chemicals. This
determines the readiness for use of the instrument.
Consumables
The following consumables are required for the operation of the
TresCon NOx-N:
z
10 l dist. H2O
Caution
Only use distilled water without any contamination (softener orsimilar)!
Dist. H2O is also used for zero point calibration. Contaminated dist.
H2O causes incorrect measurement results.
z
1.5 l cleaning solution
Cleaning solution causes severe corrosion
Do not allow this substance to get into the hands of children. Wear
suitable protective gloves and goggles/face mask when working with
this substance.
If this substance gets into the eyes, rinse thoroughly with water and
consult a doctor.
Warning
Only trained personnel is allowed to work on the analyzing section.
Always follow the warning instructions on the chemical containers and
the safety datasheets!
Always use protective equipment!
Duration
Approx. 1 minute
Note
You cannot quit the procedure. When the next measurement is started,
an AutoClean cleaning procedure and a calibration are triggered
automatically.
Step
Instrument status/action
1
Rinsing cleaning solution
2
Rinsing dist. H2O
Wear protective
goggles
Wear protective
gloves
Wear protective
clothing
Program sequence