NOx-N
Maintenance
3 - 3
ba43110e05
01/2011
3.1
Starting the AutoClean procedure
You can trigger an exceptional cleaning procedure. This AutoClean
procedure has no effect on the automatic AutoClean procedures.
The cleaning procedure consists of several individual steps that are
listed in the following overview:
Duration
Approx. 14 minutes
Note
You can quit the AutoClean procedure with Stop. At the end of the
cleaning procedure, a calibration is automatically triggered. After this
procedure, the instrument switches into the measuring mode.
Step
Instrument status/action
1
Rinsing cleaning solution
2
Flushing back sample
3
Rinsing cleaning solution
4
Wait
5
Rinsing sample
6
Rinsing dist. H2O
7
Wait
8
Rinsing dist. H2O
Program sequence
1
Measuring mode
Select the NOx-N measuring module with
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Start AutoClean
menu item with
68
and
confirm with
3
. The procedure begins immediately.
Rinse
cleaning solution
appears in the measured value display.
5
At the end of the procedure, an AutoCal calibration is
automatically triggered. Afterwards, the TresCon switches into
measuring mode.
Starting the AutoClean
procedure