NOx-N
Operation
2 - 5
ba43110e05
01/2011
2.2
Triggering a zero point calibration (AutoCal)
The individual steps required for a zero-point calibration run
automatically and are designated by AutoCal. The TresCon NOx-N
performs AutoCal at intervals. You can select the times and intervals in
the
AutoCal
menu, see section 2.5 D
Duration
Approx. 1 ½ minutes
Step
Instrument status/action
1
Rinsing dist. H2O
2
Feeding H2O dest
3
Wait
4
Measuring the E1H2O and E2H2O absorbance
Calibration
1
Measuring mode
Select TresCon NO
X
-N with
68
.
2
Press the key,
1
. The AutoCal procedure begins
immediately. Firstly,
Rinse dist. H2O
appears on the display.
Starting AutoCal