11 Correction functions
→
In the correction functions menu, call up the refer-
ence absorbance menu item.
The following display appears:
The reference absorbance function is active and
appears in the special functions menu.
→
To measure the reference absorbance, call up the
special functions submenu in the setup menu.
reference absorbance function:
→
Select the menu item, On
→
Confirm with
.
→
Call up the reference absorbance menu item.
→
Call up the menu item, measure ref. absorb.
11.2
Reference absorbance
Each absorbance measurement is made against the
basic absorbance stored in the meter.
When measuring the reference absorbance, this value
is ignored and the value measured as the reference
absorbance is used instead.
The reference absorbance function is not active when
delivered.
The measured reference absorbance value remains
stored until
●
the photometer is switched off
●
the cell type (10mm, 20mm, 50mm, round) is
changed
●
the wavelength is changed
●
the measuring mode, reference absorbance, is left
●
it is manually erased using the erase ref. absorb.
function
The menu item, erase ref. absorb. first ap-
pears after at least one reference absor-
bance has been measured.
Summary of Contents for PhotoLab Spektral
Page 2: ......
Page 32: ......
Page 38: ...8 List of contents...
Page 128: ...98 13 User defined methods...
Page 132: ...102 14 Photometer setup...
Page 140: ...110 16 RS 232 C interface...
Page 144: ...114 18 Exchanging the Memory Card...
Page 152: ...122 21 Appendix 21 1 Menu structure of the setup menu...
Page 153: ...123 21 Appendix...
Page 154: ...124 21 Appendix...
Page 160: ......
Page 161: ......