Operation
pHotoFlex / pHotoFlex Turb
80
ba75491e05
04/2009
4.13
Administrating user-defined methods
User-defined programs can be
z
entered
z
read out
z
erased.
To store user-defined programs (methods) in the meter, determine a
calibration line for your method yourself and transmit the data of this
calibration line to the pHotoFlex (Turb).
Administration of the method data can be carried out in the LSdata soft-
ware (see operating manual LSdata) or with the aid of a terminal pro-
gram (section 4.13.1).
You need the following data in any case:
Data
Possible entries/examples
Program No:
900 ... 999
Model:
Any name
(max. 5 characters)
Cuvette [mm]:
16 or 28
Wave length [nm]:
436, 517, 557, 594, 610 or 690
MRB:
(measuring range beginning)
e.g. 0.05
MRE:
(measuring range end)
e.g. 8.00
E0: (Offset)
e.g. 0.0
Slope: e.g.
1.0
Resolution:
0.0001, 0.001, 0.01, 0.1, 1, 2, 5, 10
or 100
Formula: (citation form)
e.g. PO4-P (max. 9 characters)
Unit
e.g. mg/l (max. 7 characters)
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...