Operation
pHotoFlex / pHotoFlex Turb
58
ba75491e05
04/2009
Preparatory activities
Perform the following preparatory activities when you want to calibrate:
4.6.6
Carrying out the TEC and NIST/DIN calibration procedures
The two calibration procedures only differ in the usage of different buf-
fer sets (see section 4.6.5). Make sure the Cal. type is correctly set in
the pH & ORP / Calibration menu (see section 4.6.4).
For this procedure, use any one, two or three WTW technical buffer so-
lutions in ascending or descending order.
The TEC calibration is described below. With the NIST/DIN calibration,
the NIST/DIN buffer recognition and different nominal buffer values are
displayed. Apart from that, the procedure is identical.
Note
The TEC calibration for pH 10.01 is optimized for the WTW technical
buffer solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions
can lead to an erroneous calibration. The correct buffer solutions are
given in the WTW catalog or on the Internet.
1
Connect the pH electrode to the meter.
The pH measured value display is displayed on the screen.
2
Keep the buffer solutions ready. Adjust the temperature of the
buffer solutions, or measure the current temperature if you
measure without a temperature sensor.
1
Press the <M> key (short pressure) repeatedly until the mea-
sured parameter, pH or ORP is selected.
2
Start the calibration with <CAL/ZERO>.
The calibration display appears.
3
Immerse the electrode in buffer solution 1.
i
Buffer recognition TEC
i
Immerse sensor in buffer 1
Continue
pH & ORP \ Calibration
Summary of Contents for pHotoFlex
Page 1: ...ba75491e05 04 2009 pHotoFlex pHotoFlex Turb Operating manual Handheld photometer...
Page 6: ...Contents pHotoFlex pHotoFlex Turb 6 ba75491e05 04 2009...
Page 12: ...Overview pHotoFlex pHotoFlex Turb 12 ba75491e05 04 2009...
Page 16: ...Safety pHotoFlex pHotoFlex Turb 16 ba75491e05 04 2009...
Page 22: ...Commissioning pHotoFlex pHotoFlex Turb 22 ba75491e05 04 2009...
Page 84: ...Operation pHotoFlex pHotoFlex Turb 84 ba75491e05 04 2009...
Page 100: ...Accessories options pHotoFlex pHotoFlex Turb 100 ba75491e05 04 2009...
Page 106: ...Lists pHotoFlex pHotoFlex Turb 106 ba75491e05 04 2009...