Ion concentration
pH/ION 7320
52
ba75935e04
12/2017
Calibration
evaluation
After calibrating, the meter automatically evaluates the calibration.
Calibration record
(example)
Menu item
Setting/func-
tion
Description
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration record.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<F2>
/
[USB output]
.
Quit the display with
<F1>
/
[Back]
or
<MENU/ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
-
Outputs the calibration records to
the interface.
Display
Calibration record
Magnitude of the slope [mV]
+++
50.0 ... 70.0 *
or 25.0 ... 35.0 **
Error
Error
< 30 or > 90.0 *
or < 15 or > 45 **
Error elimination (see section 14.2
ISE, page 111)
*
monovalent ions
** divalent ions
pH/ION 7320
Ser. no. 12345678
CALIBRATION ISE
18.01.2013 08:09:10
Sensor
Ser. no. 12345678
Standard 1
0.010 mg/l
Standard 2
0.020 mg/l
Voltage 1
38.5 mV
Voltage 2
58.0 mV
Temperature 1
24.0 øC
Temperature 2
24.0 øC
Ion type
Ag
Slope
64.7 mV
Sensor
+++
_____________________________________