Settings
pH 7310
44
ba75929e02
10/2011
7.3.2
Data storage
This menu contains all functions to display, edit and erase stored measured
values and calibration records (see section 8 D
7.4
Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
7.4.1 Resetting the measurement settings
The following settings for pH measurements are reset to the default settings
with the
Reset
function:
System
/
Interface
/
Output header
Output of a header for
Output
format
:
CSV
System
/
Printer
Settings for the integrated prin-
ter of the pH 7310 P.
For details, see section 10
P
System
/
Clock
Date format
Date
Time
Time and date settings (see
section 4.4.6 E
System
/
Service
information
Hardware version and software
version of the meter are dis-
played.
System
/
Reset
-
Resets the system settings to
the delivery condition (see sec-
tion 7.4.2 R
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer
AutoCal
TEC
Cal. interval
7 d
Unit for slope
mV/pH
Menu item
Possible set-
ting
Description