pH 3300i/3400i
Operation
29
ba75682e02
03/2008
4.3.3
ConCal
Single-point
calibration
Use any buffer solution for this rapid method.
The calibration will be the more exact the nearer the pH value of the buffer
solution is to that of the test sample.
Two-point
calibration
For this procedure, use two buffer solutions:
z
pH 7.0 ± 0.5
z
any other buffer solution
Note
Skip the steps 2 and 9 if you use a pH electrode with a temperature sensor.
Note
At this point, the ConCal calibration can be broken off with
<M>
. This corre-
sponds to a
single-point calibration
. For this, the instrument uses the
1
Press the
<CAL>
key repeatedly until the display
ASY
and the func-
tion display
ConCal
appears. The sensor symbol displays the evalu-
ation of the last calibration (or no sensor symbol in the delivery state
or after the measurement parameter has been reset).
2
If necessary, enter the temperature of the first buffer solution with
<
▲
> <
▼
>
.
3
Immerse the pH electrode in the first buffer solution (pH 7.0 ± 0.5 in
two-point calibration).
4
Press the
<RUN/ENTER>
key.
The measured pH value appears on the display.
5
Set the measured value to the nominal pH value of the buffer solution
(at the current temperature) with
<
▲
> <
▼
>
.
6
Press the
<RUN/ENTER>
key. The value of the asymmetry (mV) and
the sensor symbol appear on the display.
7
Press the
<RUN/ENTER>
key.
SLO
appears on the display.